Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage. 2 • 35-3010A-13 Sample-Draw Detector...
Page 3
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
Oxygen: ± 0.5% O WARNING: When using the 35-3010A-13, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 35-3010A-13 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 35-3010A- 13 as described in this manual.
The housing door is hinged on the left side and is secured by two latches on the right side. Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet 6 • 35-3010A-13 Sample-Draw Detector...
Replace the filter when it appears dirty, discolored, or clogged. If a liquid other than water is drawn into the filter, replace the filter as soon as possible. 35-3010A-13 Sample-Draw Detector • 7...
Page 8
S channels (if installed). This causes a downscale reading at the monitor on these channels. The low flow level is factory-set at 0.6 SCFH (±0.2 SCFH). NOTE: There is no low flow indication for this version of the 35-3010. 8 • 35-3010A-13 Sample-Draw Detector...
The sensor connector mates to a 7-position socket that is wired to the main PCB. Carbon Monoxide Sensor The CO sensor position in the flow block is occupied by a plastic dummy plug in the 35-3010A-13. Hydrogen Sulfide Gas Sensor The H S sensor position in the flow block is occupied by a plastic dummy plug in the 35-3010A-13.
Page 10
The test points are located on the left side of the transmitter and are labeled TP+ and TP-. A 100 mV - 500 mV output is available at these test points for use during calibration. 10 • 35-3010A-13 Sample-Draw Detector...
Page 11
The test points are located on the top of the transmitter on either side of the transmitter interconnect terminal strip. They are labeled TP+ and TP-. A 100 mV - 500 mV output is available at these test points for use during calibration. 35-3010A-13 Sample-Draw Detector • 11...
Page 12
CO sensor signal and the one on the right is for the H S sensor signal. Since the CO and H S sensors are replaced with a dummy plug in the 35-3010A-13 the CO and H S sensor signal cables carry no signal in this version of the 35-3010.
Page 13
4-20 mA signal from the CO and H S channels which will cause a downscale reading at the controller or recording device. The 35-3010A-13 does not have a CO or H sensor installed so the relay has no effect in this version of the 35-3010.
Tube, 2X Figure 6: Outline and Mounting Dimensions 2. Close and latch the housing door. NOTE: The sample-draw detector is shipped with the mounting feet “tucked under” the housing to protect the mounting feet during shipment. 14 • 35-3010A-13 Sample-Draw Detector...
7. Route the cable or wires in conduit leading from the sample-draw detector to the controller. 8. Connect the drain wire to an available chassis ground at the controller end. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire. 35-3010A-13 Sample-Draw Detector • 15...
Page 16
4/20 F ee db ack LEL (C on trolle r 24 VD C (Controller Transmitter Tra nsm itter 24 VD C Te rm inals) 4-20 ( Fe edba ck) Te rm inals) Figure 7: External (Field) Wiring, Sample-Draw Detector 16 • 35-3010A-13 Sample-Draw Detector...
Page 17
O xy gen S ensor CO and H2S S ensors Plug I nto (N ot (Not F ar S ide of P reamp Used) Used) Circuit B oard Figure 8: Internal (Factory) Wiring, Sample-Draw Detector 35-3010A-13 Sample-Draw Detector • 17...
: Verify that the sample-draw detector is sampling a fresh air environment (environment known to be free of combustible gas and of normal oxygen content, 20.9%). 2. Open the housing door. 3. Set a voltmeter to measure in the millivolt (mV) range. 18 • 35-3010A-13 Sample-Draw Detector...
A flow reduction can be caused by a flow blockage, a leak in the flow system, a malfunctioning pressure switch, or a malfunctioning pump. NOTE: There is no flow fail indication for this version of the 35-3010. 35-3010A-13 Sample-Draw Detector • 19...
1. At the sample-draw detector, set the correct flow rate with the bypass valve or flow adjust potentiometer. 2. If you cannot set the correct flow rate, check the sample lines for obstructions or kinks. 20 • 35-3010A-13 Sample-Draw Detector...
5. If the fail condition continues, replace the sensor from the problem channel or channels as described later in this section. 6. If the fail condition continues, contact RKI Instruments, Inc. for further instruction. Slow or No Response/Difficult or Unable to Calibrate Symptoms •...
6. Place the new hydrophobic filter back into the metal clamp. Replacing the Charcoal Filter NOTE: The charcoal filter does not normally need to be replaced in the 35-3010A-13 because it does not include a CO sensor. 1. Turn off or unplug power to the controller.
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 9 to 12 months is adequate for most sample draw detector applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 9 months for the sample draw detector.
Preparing for Calibration section above. 5. If necessary, use a small flat-blade screwdriver to adjust the zero pot for the H channel until the voltmeter reading is 100 mV ± 2 mV. 24 • 35-3010A-13 Sample-Draw Detector...
3. Wait 1 to 2 minutes to allow the calibration gas to be drawn out and the reading to stabilize. 4. Close the housing door. 5. Store the components of the calibration kit in a safe and convenient place. 35-3010A-13 Sample-Draw Detector • 25...
1. Screw the regulator into the zero air calibration cylinder. 2. Connect the calibration kit sample tubing to the regulator. 3. Connect the sample tubing from the regulator to the inlet line at or near the INLET fitting. 26 • 35-3010A-13 Sample-Draw Detector...