User Values For Derived Parameters - Xylem ODO RTU User Manual

Optical dissolved oxygen & temperature sensor
Table of Contents

Advertisement

5.5

User Values for Derived Parameters

Several values reported by the probe are dependent upon parameters that the probe does not supply. It is, therefore, a
requirement of the user to supply these values if the derived parameters are of interest.
Salinity
Salinity values are not required (or available to be set) for ODO/CT probes, as the parameter value is already available. In ODO/T
probes, salinity is required for ODO mg/L calculations. The salinity value is set by writing an entire block of values (non-divisible)
using a write multiple registers command (function code 0x10). This information can be broadcast to all devices on the network and
updated periodically. The default on power-up is 0.0.
Address
Description
0x0400
User Salinity Value (ppt) [3:0]
0x0401
Request
Function Code
Register Address
Number Registers
Byte Count
Register Values
Water Pressure
In ODO/CT probes, water pressure is required for salinity calculations. The water pressure value is set by writing an entire block
of values (non-divisible) using a write multiple registers command (function code 0x10). This information can be broadcast to all
devices on the network and updated periodically. The default on power-up is 0.0.
Address
Description
0x0402
User Water Pressure Value (psi) [3:0]
0x0403
Request
Function Code
Register Address
Number Registers
Byte Count
Register Values
Modbus Configuration & Commands
1 byte
2 bytes
2 bytes
1 byte
4 bytes
1 byte
2 bytes
2 bytes
1 byte
4 bytes
0x10
0x0400
0x0002
0x04
User Salinity Value (ppt) [3:0]
0x10
0x0402
0x0002
0x04
User Pressure Value (psi) [3:0]
26
R/W
Type
Default
RW
Float
0.0
R/W
Type
Default
RW
Float
0.0