Xylem ODO RTU User Manual page 23

Optical dissolved oxygen & temperature sensor
Table of Contents

Advertisement

Conductivity
Factory Reset
ODO values can be reset to the factory calibration by issuing a write single register command (function code 0x06) at address
0x0200 with a value of 0x0001.
Request
Function Code
Register Address
Register Value
Last Calibration Time and QC Score
The time of the last conductivity calibration and QC score can be queried to help track internal processes for keeping the devices
properly calibrated. The last time must be maintained during calibrations by the user. These registers do not support writes, as time
is written in each calibration command. This is only intended to be read. Last calibration time is a 32-bit (two register) value, and
is recommended to be set as epoch time. The last calibration time and QC score are available as read-only using a read holding
registers command (function code 0x03) starting at address 0x0310.
QC Score
Best
OK
Bad
Address
Description
0x0310
Conductivity Last Calibration Time (s) [3:0]
0x0311
0x0312
Conductivity QC Score
Request
Function Code
Register Address
Number Registers
Response
Function Code
Byte Count
Register Values
Modbus Configuration & Commands
1 byte
2 bytes
2 bytes
1 byte
2 bytes
2 bytes
1 byte
1 byte
6 bytes
0x06
0x0300
0x0001
0x00
0x01
0x02
0x03
0x0310
0x0003
0x03
0x06
Conductivity Last Calibration Time (s) [3:0]
Conductivity QC Score
23
R/W
Type
Default
R
U32
N/A
R
U16
N/A