Xylem ODO RTU User Manual page 21

Optical dissolved oxygen & temperature sensor
Table of Contents

Advertisement

ODO Cap Replacement
When replacing an ODO cap, new coefficients are required. These values will be included with the cap replacement. The
coefficients are written in an entire block of values (non-divisible) using a rite multiple registers command (function code 0x10).
ODO Cap Replacement Time (s) is tracked by the user, and is suggested to hold an epoch timestamp for the operation.
Address
Description
0x0100
ODO Cap Coefficient K1 [3:0]
0x0101
0x0102
ODO Cap Coefficient K2 [3:0]
0x0103
0x0104
ODO Cap Coefficient K3 [3:0]
0x0105
0x0106
ODO Cap Coefficient K4 [3:0]
0x0107
0x0108
ODO Cap Coefficient K5 [3:0]
0x0109
0x010A
ODO Cap Coefficient K6 [3:0]
0x010B
0x010C
ODO Cap Coefficient K7 [3:0]
0x010D
0x010E
ODO Cap Coefficient KC [3:0]
0x010F
ODO Cap Replacement Time (s) [3:0]
0x0110
Request
Function Code
Register Address
Number Registers
Byte Count
Register Values
Modbus Configuration & Commands
1 byte
2 bytes
2 bytes
1 byte
34 bytes
0x10
0x0100
0x0011
0x22
ODO Cap Coefficient K1 [3:0]
ODO Cap Coefficient K2 [3:0]
ODO Cap Coefficient K3 [3:0]
ODO Cap Coefficient K4 [3:0]
ODO Cap Coefficient K5 [3:0]
ODO Cap Coefficient K6 [3:0]
ODO Cap Coefficient K7 [3:0]
ODO Cap Coefficient KC
ODO Cap Replacement Time (s) [3:0]
21
R/W
Type
Default
RW
U32
N/A
RW
U32
N/A
RW
U32
N/A
RW
U32
N/A
RW
U32
N/A
RW
U32
N/A
RW
U32
N/A
RW
U16
N/A
RW
U32
N/A

Advertisement