[threshold] – Trigger threshold in A/D counts.
Examples
We have a strobe attached so that when the vibration exceeds 0.25 ips the strobe will
activate. To do this we program tier 1 for 0.25 ips. To determine the A/D count we use the
last equation shown for the TS command. We are assuming a 10.24 ips range.
TT 1 1 1 33568 is the command.
Later we add a horn so that when the vibration exceeds 0.5 ips the horn will sound. To do
this we program tier 2 for 0.5 ips. To determine the A/D count we use the last equation
shown for the TS command. We are assuming a 10.24 ips range.
TT 1 1 2 34368 is the command.
Mini-Seis III Pro™ Operating Manual
0.25
(
∗ 32768) + 32768 = 33568
10.24
0.5
(
∗ 32768) + 32768 = 34368
10.24
Page 48