Command Description
SSP2 4 unit
Used to set single channel parameters for channel 4. It is also
max min
used to get or set sensor full-scale range step size for a Unit
type. If passed with only a channel, returns the current values
for that channel in the format:
<channel as integer> <unit as integer> <max as float> <min as
float>
TA raw
Get or set acoustic trigger. If passed with no parameters,
returns the current value.
The threshold in units is given by:
threshold_units=range•(raw-32768)/32768
The threshold in counts is given by:
threshold_counts=raw-32768
As an example, if the range is 5.12 mb, valid values of raw are
32784 through 65520. If raw is 32784, the threshold in units is
0.0025 mb, and the threshold in counts is 16.
TM onEn
Enable/disable and set start and stop timers. If the start and
onHH onMM
stop times are the same, the start timer is enabled, but the stop
offEn offHH
timer is disabled. If sent with no parameters, returns the current
offMM
settings.
Mini-Seis III Pro™ Operating Manual
Parameters
The unit-type is ignored
for A-weighted air and is
always dBA.
Min is only used for A-
weighting and is the
minimum reading (in dBA)
for the channel.
Channel is the index of the
single channel to
configure. The single
channel must be channel
4.
Unit is the displayed value
(and also specifies the
type of measurement,
e.g., seismic or air). Valid
values are:
 0: Disabled
 1: IPS (seismic)
 2: MMPS (seismic)
 3: dBL (air)
 4: mb (air)
 5: Pa (air)
 6: PSI (air)
 7: dBA (air with A-
weighting card)
All other values are
reserved for future use.
Max is the maximum
reading for the channel.
Min is the minimum
reading for the channel.
raw:
The raw ADC reading at
which to trigger, 32784–
65520 in 16-count
increments.
onEn and offEn enable or
disable the start and stop
timer, respectively.
0 = disable
1 = enable
onHH and onMM are the
hour and minute at which
to turn on data collection,
and offHH and offMM are
the hour and minute at
which to turn off data
collection.
HH: 00–23
MM: 00–59
Page 42