Operating manual
Anemometer
HD2303.0
www.deltaohm.com
Keep for future reference.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Delta OHM HD2303.0 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GHM Delta OHM HD2303.0

  • Page 1 English Operating manual Anemometer HD2303.0 Companies / Brands of GHM www.deltaohm.com Keep for future reference.
  • Page 2: Table Of Contents

    CONTENTS GENERAL CHARACTERISTICS ..........................3 DESCRIPTION OF THE FUNCTIONS ........................6 THE PROGRAMMING MENU ............................. 9 PROBES AND MEASUREMENTS ..........................10 ............................10 IND SPEED MEASUREMENT Flow rate measurement ............................11 SICRAM ..........12 WIRE WIND SPEED MEASUREMENT PROBES COMPLETE WITH MODULE Operation .................................
  • Page 3: General Characteristics

    GENERAL CHARACTERISTICS The Anemometer Model HD2303.0 is a portable instrument that allows to perform measurements in the fields of air conditioning, heating and ventilation. It is fitted with a large LCD display for excellent visualization of the measured data. The Anemometer Model HD2303.0: •...
  • Page 4 Anemometer HD2303.0 - 4 - HD2303 V1.6...
  • Page 5 HD2303.0 1. Input for probes, 8-pole DIN45326 connector. 2. Battery symbol: displays the battery charge level. 3. Function indicators. 4. Secondary display line. 5. DATA/ENTER key: during normal operation displays the maximum (MAX), the minimum (MIN) and the average (AVG) of current measurements; in the menu, confirms the current selection.
  • Page 6: Description Of The Functions

    DESCRIPTION OF THE FUNCTIONS The keyboard of the Anemometer Model HD2303.0 is composed of double-function keys. The function on the key is the "main function", while the one above the key is the "secondary function". When the instrument is in standard measurement mode, the main function is active. Once the Settings Menu has been opened, by pressing simultaneously the DATA+UNIT keys, the secondary function is enabled.
  • Page 7 CLR/ESC key The "CLR" key has two functions: • CLEAR (CLR): allows to reset the maximum (MAX), minimum (MIN) and average (AVG) value of the captured measurements; • ESC: once the MENU has been opened with the DATA+UNIT keys, the CLR key will allow to cancel the parameters set using the ...
  • Page 8 UNIT/MENU key The UNIT key is used for the following functions: • UNIT: by pressing this key the unit of measurement of the main input quantity is selected: the unit of measurement will appear in the upper part of the display; the measured value will be displayed in the central line.
  • Page 9: The Programming Menu

    THE PROGRAMMING MENU To access to the menu press simultaneously the following keys: The items to be set are listed in this order: 1. SECT m - SECT inch : this parameter defines a duct's section area for flow rate calculation.
  • Page 10: Probes And Measurements

    PROBES AND MEASUREMENTS The Anemometer works with hot-wire, vane and temperature probes fitted with the SICRAM module. It works also with temperature probes with 4 wire Pt100 or 2 wire Pt1000 sensor. Some probes are fitted with SICRAM module that acts as an interface between the sensor on the probe and the instrument.
  • Page 11: Flow Rate Measurement

    Flow rate measurement The flow rate measurement requires knowledge of the duct or vent area orthogonal to the flow: the menu items indicated by "SECT m2" and "SECT INC2" define the section area m or inch To set the area value: press simultaneously DATA and UNIT to open the menu;...
  • Page 12: Hot - Wire Wind Speed Measurement Probes Complete With Sicram Module

    SICRAM WIRE WIND SPEED MEASUREMENT PROBES COMPLETE WITH MODULE The hot-wire wind speed measurement probes, fitted with SICRAM module, are the models: AP471S1, AP471S2, AP471S3, AP471 S4 The AP471S1 and AP471S3 probes measure incident air flows up to 40m/s. The AP471S2, AP471S4 probes are fitted with an omnidirectional sensor allowing measurement of speeds up to 5m/s in any direction of the air flow incident on the probe.
  • Page 13: Operation

    Operation Where present, extend the telescopic rod to the necessary length paying attention to the cable so that it can slide freely and without strain. Uncover the sensor and introduce the probe in the air flow being measured, maintaining the arrow at the top of the probe parallel to the flow, as indicated in the figures.
  • Page 14: Dimensions

    Dimensions 200 min - 880 max Ø8 Ø13 AP471 S1 Ø8 Ø13 AP471 S2 200 min 880 max Ø8 Ø13 AP471 S3 - 14 - HD2303 V1.6...
  • Page 15 AP471S4 - 15 - HD2303 V1.6...
  • Page 16: Vane Wind Speed Measurement Probes Complete With Sicram Module

    SICRAM ANE WIND SPEED MEASUREMENT PROBES COMPLETE WITH MODULE The AP472S1 and AP472S2 vane probes measure the incident wind speed and flow rate. The probe AP472S1 measures also the temperature using a thermocouple of type K. On request, they can be fitted with a telescopic rod that eases measurements in areas difficult to reach (for example vents).
  • Page 17: Care And Maintenance Of The Probes

    The probe should be maintained orthogonal to the flow and not tilted in relation to it The probe is correctly positioned in relation to the air-flow when the value measured is the maximum possible. Proceed with measurement following the instructions provided in this chapter. Care and maintenance of the probes The probe performance, mainly at low speeds, largely depends on the very slow friction of the vane on its own axis.
  • Page 18: Dimensions

    Dimensions AP472S1 AP472S2 To mount the telescopic rod on the probes, unscrew the handle (3) holding the probe body still in the point (1). Screw the rod end on the screw (2). The handle (3) is screwed on the telescopic rod. - 18 - HD2303 V1.6...
  • Page 19: Direct Input Pt100 And Pt1000 Temperature Probes

    Pt100 Pt1000 T IRECT INPUT EMPERATURE PROBES The instrument accepts the input of Platinum temperature probes with resistances of 100Ω (Pt100) and 1000Ω (Pt1000). The Pt100 are connected to 4 wires, the Pt1000 to 2 wires; the excitation current was chosen in order to minimize the sensor self-heating effects.
  • Page 20: Direct Connection Of 4-Wire Pt100 Sensors

    5. open the two module shells: the printed circuit to which the probe must be connected is housed inside. On the left there are the 1…4 points on which the sensor wires must be welded. The JP1…JP4 jumpers are in the center of the board. These must be closed with a tin bead for some type of sensors: Not Used Ni1000...
  • Page 21: Warnings And Operating Instructions

    WARNINGS AND OPERATING INSTRUCTIONS 1. Temperature probes are not insulated from their external casing; be very careful not to come into contact with live parts (above 48V). This could be extremely dangerous for the instrument as well as for the operator, who could be electrocuted. 2.
  • Page 22: Instrument Signals And Faults

    INSTRUMENT SIGNALS AND FAULTS The following table lists all error indications and information displayed by the instrument and supplied to the user in different operating situations: Display indications Explanation This appears in the display central line when a temperature only probe is connected. - - - In the lower line the temperature is shown correctly.
  • Page 23: Low Battery Warning And Battery Replacement

    LOW BATTERY WARNING AND BATTERY REPLACEMENT The battery symbol on the display constantly shows the battery charge status. To the extent that batteries have discharged, the symbol "empties". When the charge decreases still further it starts blinking. In this case, batteries should be replaced as soon as possible. If you continue to use it, the instrument can no longer ensure correct measurement.
  • Page 24: Notes About Working And Operative Safety

    NOTES ABOUT WORKING AND OPERATIVE SAFETY Authorized use The technical specifications as given in chapter TECHNICAL CHARACTERISTICS must be observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized. Any other use is considered unauthorized. General safety instructions This measuring system is constructed and tested in compliance with the EN 61010-1 safety regulations for electronic measuring instruments.
  • Page 25: Instrument Technical Characteristics

    INSTRUMENT TECHNICAL CHARACTERISTICS ECHNICAL INFORMATION ON THE NEMOMETER Instrument Dimensions (Length x Width x Height) 140 x 88 x 38 mm Weight 160 g (complete with batteries) Material Display 2x4½ digits plus symbols Visible area: 52 x 42 mm Operating conditions -5 ÷...
  • Page 26: Technical Data Of Probes And Modules In Line With The Instrument

    ECHNICAL DATA OF PROBES AND MODULES IN LINE WITH THE INSTRUMENT IND SPEED MEASUREMENT PROBES Hot-wire probes: AP471S1 - AP471S2 - AP471S3 - AP471S4 AP471S1 - AP471S3 AP471S2 AP471S4 Type of measure Wind speed, calculated flow rate, air temperature Type of sensor Speed NTC thermistor Omnidirectional NTC thermistor...
  • Page 27 Vane probes: AP472S1 - AP472S2 AP472S1 AP472S2 Wind speed, Wind speed, Type of measure calculated flow rate, calculated flow rate air temperature Diameter 100mm 60mm Type of measurement Speed Vane Vane Temperature K thermocouple ---- Measurement range Speed 0.6…25 m/s 0.5…20 m/s Temperature -25…+80 (*)
  • Page 28: Emperature Probes Pt100 Sensor Using Sicram Module

    Pt100 EMPERATURE PROBES SENSOR USING SICRAM MODULE Model Type Application range Accuracy TP472I Immersion -196°C…+500°C TP472I.O Immersion -50°C…+300°C TP473P.I Penetration -50°C…+400°C Penetration -50°C…+300°C TP473P.O TP474C.O Contact -50°C…+300°C TP475A.O -50°C…+250°C ±0.1°C (@ 0 °C) TP472I.5 Immersion -50°C…+400°C ±0.2°C (-50 °C ≤ t ≤ 250 °C) Immersion -50°C…+400°C TP472I.10...
  • Page 29: Order Codes

    ORDER CODES HD2303.0 The kit is composed of the HD2303.0, 3 1.5V alkaline batteries, operating manual, and case. The probes must be ordered separately. SICRAM ROBES COMPLETE WITH MODULE WIND SPEED MEASUREMENT PROBES HOT-WIRE PROBES: AP471S1 Hot-wire telescopic probe, measuring range: 0.1…40 m/s. Cable length 2 metres. AP471S2 Omnidirectional hot-wire telescopic probe, measuring range: 0.1…5 m/s.
  • Page 30 TP472I.10 Immersion probe, sensor Pt100. Stem Ø6 mm, length 1000 mm. Cable length 2 metres. TP49A.I Immersion probe, sensor Pt100. Stem Ø2.7 mm, length 150 mm. Cable length 2 metres. Aluminium handle. TP49AC.I Contact probe, sensor Pt100. Stem Ø4 mm, length 150 mm. Cable length 2 metres.
  • Page 32 G U A R A N T E E TERMS OF GUARANTEE All DELTA OHM instruments are subject to accurate testing, and are guaranteed for 24 months from the date of purchase. DELTA OHM will repair or replace free of charge the parts that, within the warranty period, shall be deemed non efficient according to its own judgement.
  • Page 33 GHM GROUP – Delta OHM | Delta Ohm S.r.l. a socio unico Via Marconi 5 | 35030 Caselle di Selvazzano | Padova | ITALY Phone +39 049 8977150 | Fax +39 049 635596 www.deltaohm.com | sales@deltaohm.com The quality level of our instruments is the result of the constant development of the product. This may produce some differences between the information written in this manual and the instrument you have purchased.