Analog inputs
Thermocouple
Designation
Type Standard
Fe-CuNi
L
DIN 43 710
Fe-CuNi
J
EN 60 584
Cu-CuNi
U
DIN 43 710
Cu-CuNi
T
EN 60 584
NiCr-Ni
K
EN 60 584
NiCr-CuNi
E
EN 60 584
NiCrSi-NiSi
N
EN 60 584
Pt10Rh-Pt
S
EN 60 584
Pt13Rh-Pt
R
EN 60 584
Pt30Rh-Pt6Rh B
EN 60 584
W3Re/W25Re D
W5Re/W26Re C
W3Re/W26Re
Chromel-copel
GOST R 8.585-2001
Chromel-alumel
GOST R 8.585-2001
PLII (Platinel II)
Shortest span
Range start/end
Cold junction
Cold junction accuracy (internal)
Cold junction temperature (external)
Sampling cycle
Input filter
Electrical isolation
Resolution
Features
1
The linearization accuracy refers to the maximum measuring range. The linearization accuracy is reduced with short spans.
Resistance thermometer
Designation
Standard
Pt100
EN 60 751
(TC = 3.85*10
Pt100
JIS 1604
(TC = 3.917*10
Pt100
GOST 6651-94 A.1
(TC = 3.91*10
Pt500
EN 60 751
(TC = 3.85*10
Pt1000
EN 60 751
(TC = 3.85*10
Ni 100
DIN 43 760
(TC = 6.18*10
Pt50
ST RGW 1057 1985
(TC = 3.91*10
Cu 50
(TC = 4.26*10
/

6 Technical data (extract from data sheet)

Meas. range
-200 to +900°C
-200 to +1200°C
-200 to +600°C
-270 to +400°C
-200 to +1372°C
-200 to +1000°C
-100 to +1300°C
0 to
0 to
0 to
0 to
0 to
0 to
-200 to +800°C
-200 to +1372°C
0 to
Connection circuit Meas. range
2/3-wire
-3
2/3-wire
1/°C)
4-wire
2/3-wire
-3
2/3-wire
1/°C)
4-wire
2/3-wire, 4-wire
-3
2/3-wire, 4-wire
1/°C)
2/3-wire, 4-wire
-3
2/3-wire, 4-wire
1/°C)
2/3-wire
-3
2/3-wire
1/°C)
4-wire
2/3-wire, 4-wire
-3
1/°C)
2/3-wire
-3
2/3-wire
1/°C)
4-wire
4-wire
-3
2/3-wire
1/°C)
2/3-wire
4-wire
4-wire
1768°C
1768°C
1820°C
2495°C
2320°C
2400°C
1395°C
Type L, J, U, T, K, E, N, chromel-alumel, PLII: 100°C
Type S, R, B, D, C, W3Re/W26Re, chromel-copel: 500°C
freely programmable within the limits, in 0.1°C steps
Pt100 internal or thermostat external constant
-50 to +150°C adjustable
Channel 1 - 18: 125ms in total
2nd order digital filter; filter constant adjustable from 0 to 10.0 sec
see "Electrical data" on page 5 and
"Overview of the electrical isolation" on page 16
also programmable in °F
-200 to +100°C
-200 to +850°C
-200 to +850°C
-200 to +100°C
-200 to +650°C
-200 to +650°C
-200 to +100°C
-200 to +850°C
-200 to +100°C
-200 to +850°C
-200 to +100°C
-200 to +850°C
-200 to +850°C
-60 to +180°C
-200 to +100°C
-200 to +1100°C
-200 to +100°C
-200 to +1100°C
-50 to +100°C
-50 to +200°C
-50 to +100°C
-50 to +200°C
1
Accuracy
±0.1%
±0.1% from -100°C
±0.1% from -150°C
±0.1% from -150°C
±0.1% from -80°C
±0.1% from -80°C
±0.1% from -80°C
±0.15%
±0.15%
±0.15% from 400°C
±0.15% from 500°C
±0.15% from 500°C
±0.15% from 500°C
±0.15% from -80°C
±0.1% from -80°C
±0.15%
± 1°C
>14 bit
1
Accuracy
±0.5°C
±0.8°C
±0.5°C
±0.5°C
±0.8°C
±0.5°C
±0.5°C
±0.8°C
±0.5°C
±0.9°C
±0.5°C
±0.8°C
±0.5°C
±0.4°C
±0.5°C
±0.9°C
±0.5°C
±0.6°C
±0.5°C
±0.9°C
±0.5°C
±0.7°C
Meas. curr.
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 100μA
≈ 100μA
≈ 100μA
≈ 100μA
≈ 100μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
≈ 250μA
33