QA-ES User & Service Manual QA-ES Electrosurgical Analyzer P/N 14025...
Page 2
METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the QA-ES Electrosur- gical Analyzer. Neither METRON nor its local representatives will accept, nor be bound by any other form of guarantee concerning the QA-ES Electrosurgical Analyzer other than this guarantee.
Page 3
Table of Contents 1. Introduction............................1 2. Installation............................1 3. Operating QA-ES..........................1 4. ESU Tests with QA-ES........................1 5. Control and Calibration........................1 6. Component Functions and Parts......................1 Appendix A – Diagrams ........................A-1 Appendix B – Error Report Form......................B-1 Appendix C – Improvement Suggestion Form..................C-1...
Page 5
Manual Revision Record This record page is for recording revisions to your QA-ES User & Service Manual that have been published by METRON AS or its authorized representatives. We recommend that only the man- agement or facility representative authorized to process changes and revisions to publications: •...
Page 7
(REM) test Testing is accomplished by measuring the ESU output against test loads that are set and adjusted in the QA-ES. The QA-ES can auto- matically execute a power distribution test with a load resistance ranging from 10 ohms to a maximum of 5200 ohms. The automatic...
Page 8
34.2 cm / 11.61 in. Length 39.5 cm / 13.41 in. Weight 9.8 kg / 21.6 lbs. Standard Accessories: QA-ES Electrosurgical Analyzer (P.N. 14010) Power Cord (P.N. 14300) QA-ES User and Service Manual (P.N. 14025) Additional Accessories: E-Input Measuring Cable - Black (P.N. 11451)
Page 9
(P.N. 14100) PRO-Soft QA-ES software (P.N. 12200) PRO-Soft QA-ES DEMO (P.N. 14201) PRO-Soft QA-ES User Manual (P.N. 14225) Storage: Store in the carrying case in dry surroundings within the tempera- ture range specified. There are no other storage requirements. Periodic Inspection:...
Page 11
The customer should place a new purchase order to ensure delivery. 4. When returning an instrument to METRON AS, or the company representative, fill out the address label, describe what is wrong with the instrument, and provide the model and serial numbers.
Page 12
3. If PRO-Soft QA-ES is being used, attach an RS-232 (null mo- dem/data transfer configured) cable to the 9-pin D-sub outlet port located at the rear of the QA-ES. Do not attach the printer cable to the QA-ES. See below. However, if you are not using PRO-Soft QA-ES, and are sending directly to a printer for print- outs, attach the printer cable to the 25-pin outlet port.
Page 13
NOTE test “Protocol.” The equipment data can be entered manually into PRO-Soft QA-ES has its the protocol, or it may be retrieved by PRO-Soft from a database own user manual, which program, or other equipment files.
Page 15
3. Operating QA-ES operating controls, switches This chapter explains the and menus the QA-ES, and details how to use them in ESU testing. 3.1 Control Switches and Connections Front Panel Power Switch Turns the power on and off. Encoder Sets values according to the specified range and choose between different operations/ measurement ranges.
Page 16
Right Side Panel 10. Terminal Connection for the electrode outputs of the VAR. RED and BLACK LOAD ESU. Active electrode to the red terminal, and neutral electrode to the black terminal. 11. Terminal Additional fixed load resistance of 200 ohms BLUE and BLUE FIXED LOAD 500 watt for serial connection dur- ing leakage test.
Page 17
3.2 QA-ES Menu and Function Keys The QA-ES uses a display, programmable function keys and a set- ting regulator to provide flexibility and control over the operations. The upper part of the screen displays messages, status and results. The menu bar is at the bottom of the display. The function keys are numbered from Fl to F5.
Page 18
LCD Display Menu/ Messages (Detail) 1. Startup Screen. The following screen will be displayed for 2 seconds after the QA-ES has been switched on. 2. Main Menu First Menu Bar (Page 1) b. Second Menu Bar (Page 2) 3. SHOW CHOICES (F1). This function is activated when you see an asterisk (*) in the status field under ‘Mode.’...
Page 19
Press QUIT MENU (F5) to return to the main menu. 7. PRINT HEADER (F3). Writes a heading for a new test proto- col. 8. REMOTE CONTR. (F4). Enables you to control the QA-ES through a PC. Required software: PRO-Soft QA-ES.
Page 20
This is used to trigger the foot switch input on the ESU being tested. 3.7 Main Switch On/Off The QA-ES has to be turned off for at least 5 seconds before turning it on again to allow the reset circuit to unload.
Page 21
4. ESU Tests with QA-ES This chapter explains the tests that can be conducted by the QA-ES on an ESU, as well as the features available with the PRO-Soft QA- ES software accessory. 4.1 Power Distribution This test checks the power provided by the ESU over a range of load resistances.
Page 22
Test Configuration Limits of Acceptable Leakage Current Bipolar The leakage current should not exceed 1% of the maximum bipolar rated power output. Measured at equipment The leakage current should not exceed 100 mA. terminals 1. Grounded HF Equipment: Measurements of the HF cur- rent leakage.
Page 23
Neutral electrode test setup in compliance with IEC 601.2.2, sec. 19.101b, fig, 104 and sec. 19.102. (Adopted by ANSI/AAMI HF18-1993) 4.3 REM Alarm This test ensures that the ESU will sound an alarm if the resistance between the two neutral electrodes exceeds your specified limit. The program directs the OA-ES to gradually increase the resistance.
Page 25
Afterwards, move the bottom plate of the isolation box working temperatures. by loosening the 16 screws holding it in place. Adjust the voltage switch at the rear of the QA-ES so that is corresponds to the main voltage (115 or 230V). 5.3 Function Testing 1.
Page 26
5. Foot Switch. Connect a multimeter to the foot switch output on the QA-ES (green). Check that the relay for the foot switch is connected when you press START (F3). 6. Fixed Load. Measure the value of the fixed load with a multi- meter (blue).
Page 27
Turn the wheel and press ENTER to specify a new value for Q_gain. Repeat the test until the average value between the QA-ES and URE 3 is smaller than 0.1 dB. This corresponds to an error of 2.3%. Peak Detector. Once the measuring device used in measuring the effect has been calibrated, you can calibrate the peak detec- tor.
Page 28
< 6. Scope Output. Connect the high voltage generator to 'Var. Load' and the oscilloscope to 'Scope Output'. Set the QA-ES to Remember that the continuous mode and start measuring by pressing on Start. Ac- scope output is not cali-...
Page 29
6. Component Functions and Parts This chapter provides a detailed description of the functions of the main components of the QA-ES, as well as a parts list for cross-ref- erence. Reference is made to the component location and schematic diagrams to assist servicing personnel. These diagrams are foldouts, and are located in Appendix B.
Page 30
I/O. Ul3 saves the calibration parameters for the Sam- pling Unit. The parameters can be stored independently of the processor component so that the QA-ES does not have to be re- calibrated when upgrading the software. 3. Function Keys (See QA-ES Processor Board Schematic Diagram 1).
Page 31
6.2 Sampling Unit The QA-ES Sampling Unit is placed inside the internal protection box of the instrument. The card is fixed vertically on the right side of the protection box. The unit comprises: power supply;...
Page 32
U22 does the same for the negative peak de- tector. The peak detector ran be reset by short-circuiting C27 and C28 via octol analog switch U24 (See QA-ES Processor Board Schematic Diagram 2). 6. A/D Converter...
Page 33
The measuring device and load resistances in the QA-ES are isolated from the processor unit and the housing by a galvanic shield. This protects the user if a fault in the QA-ES produces ground currents during the measurements. This interface is based upon opto-couplers and various mains transformers for the user interface and the measuring device in the QA-ES.
Page 34
V. Cl and C2 provide relay drivers U2 and U3 with +10 VDC and -10 VDC. These voltages are used for controlling the relays on the Load Board. 2. Load Resistances There are two load resistances on the Load Board, seen from the ESU being tested.
Page 35
6. Interface Towards Sampling Unit When the signals pass through an optical barrier between the processor system and the rest of the electronics, the Sampling Unit's F1 terminal controls the relays, choosing the load and at- tenuator values on the Load Board. 7.
Page 36
COMPONENT PART TYPE/VALUE QTY. DIAGRAM REFERENCE Screw recessed head poz DIN 7985 M3x10 Screw recessed head poz DIN 7985 M4x30 Nut M3 Nut M4 Safety nut M4 Spring washer M4 Washer M4 12mmø DIN 9021 Spacer for fixing the front plate Contact plate RF protection box Front plate w/5 screws for circuit board...
Page 37
COMPONENT PART TYPE/VALUE QTY. DIAGRAM REFERENCE Display-connection 20 pin Berg 71991-410 (DISP1) Board pin base 16 pin Board pin base 10 pin Board pin base 26 pin Pin base for display (DISP1) Pin base 36 pol TP1-TP7 Screw clip basic part 2 pol J4J5 Screw clip...
Page 41
Appendix A - GLOSSARY OF TERMS USED Bipolar Electrosurgery Electrosurgery where current flows between two bipolar electrodes that are posi- tioned around tissue to create a surgical effect. Current passes from one elec- trode, through the desired tissue, to the other electrode, thus completing the cir- cuit without entering any other part of the patient's body.
Page 53
Error Report Product: Version: Serial no.: Description of the situation prior to the error: Description of the error: (METRON AS internally) Comments: Received date: Correction date: Ref No. Critical Normal Minor...
Page 54
(name) Address: Fax: E-mail: Date: Improvement Suggestion Product: Version: Description of the suggested improvement: (METRON AS internally) Comments: Received date: Correction date: Ref No. Critical Normal Minor...