Micropipette manual FEATURES H series Plunger Tip ejector Volume adjustment Volume indicator H series features – Variable volume ranges from 0.5 to 1,000 µl – Universal fit with standard tips – 3-digit volume adjustment – Ergonomic design – Very low operating force –...
Page 4
Volume indicator A series features – Variable volume ranges from 0.5 to 1,000 µl – Universal fit with standard tips – 4-digit volume adjustment – High accuracy and precision – Enhanced ergonomic design – Autoclavable (121°C) up to six times www.minipcr.com...
SETTING THE VOLUME Turn the volume adjustment dial counterclockwise to increase the volume and clockwise to decrease the volume. CAUTION: Setting the volume outside of the recommended range may result in permanent damage. www.minipcr.com...
Micropipette manual CALIBRATION PROCEDURE Use these volumes when calibrating your pipettes TO CALIBRATE H SERIES PIPETTES: Insert the calibration wrench (A) into the notches found on top of the volume adjustment dial (red arrowhead in B). The wrench contains a notch (white arrow in B) that allows it to be inserted around the plunger.
Page 10
To take the pipette out of calibration mode, use the calibration tool to release the plastic latch. * If the latch doesn’t catch, check to ensure that you have successfully rotated the pointed tip of the latch all the way to the left. www.minipcr.com...