National Instruments NI-DAQ mx Calibration Procedure page 11

Ao waveform
Table of Contents

Advertisement

NI-DAQ Function Call
Call
DAQmxCfgImplicitTiming
with the following parameters:
taskHandle:
taskHandle
sampleMode:
DAQmx_Val_ContSamps
sampsPerChan:
10000
NI-DAQ Function Call
Call
DAQmxStartTask
the following parameter:
taskHandle:
taskHandle
NI-DAQ Function Call
Call
DAQmxClearTask
parameter:
taskHandle:
taskHandle
© National Instruments Corporation
4.
Configure the counter for continuous square wave generation using
DAQmxCfgImplicitTiming
5.
Start the generation of a square wave using
(DAQmx Start Task VI).
with
6.
The device will begin to generate a 5 MHz square wave when the
DAQmxStartTask
read by your counter to the test limits shown on the device table. If the
value falls between these limits, the test is considered to have passed.
7.
Clear the generation using
Task VI).
with the following
8.
Disconnect the counter from your device.
You have verified the counter on your device.
(DAQmx Timing VI).
LabVIEW Block Diagram
LabVIEW Block Diagram
function completes execution. Compare the value
DAQmxClearTask
LabVIEW Block Diagram
11
AO Waveform Calibration Procedure for NI-DAQmx
DAQmxStartTask
(DAQmx Clear