Operation

Eliwell Control Panel
Control Button
Control Button
Basic Settings
Adjusting the Set Point:
1. Switch on Mains power – the display will flash several times, then the fridge
system will switch on and the display will give a steady reading, this is the
temperature of the water bath sensor.
2. Press and release the 'Set' button – The display will show 'SET'.
3. Press 'Set' again and the display will show a numeric value (e.g. 9°C).
4. Raise or lower this figure to the desired setting using the up or down arrows
on the left of the control display window.
5. When the correct setting is shown in the display window press 'Set' button
the display will now show SET.
6. Press FNC button to return to the probe temperature reading.
Digital Display
Function Button
Setting Button