OCULUS SDI 4 Instruction Manual

OCULUS SDI 4 Instruction Manual

Stereoscopic diagonal inverter
Hide thumbs Also See for SDI 4:

Advertisement

Quick Links

®
OCULUS SDI
4
INSTRUCTION MANUAL
Stereoscopic Diagonal Inverter

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SDI 4 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for OCULUS SDI 4

  • Page 1 ® OCULUS SDI INSTRUCTION MANUAL Stereoscopic Diagonal Inverter...
  • Page 2 Our service team will gladly assist. OCULUS Optikgeräte GmbH OCULUS is certified according to EN ISO 13485, setting high standards of quality where development, manufacture, quality assurance and service regarding the entire range of products are concerned.
  • Page 3: Table Of Contents

    Table of Contents Table of Contents Scope of Delivery.........................1 Graphic Symbols SDI® 4......................2 Safety Instructions........................3 About this Manual ......................3 3.1.1 Used Graphic Symbols.................3 Safety Instructions for Use ..................4 Proper Use .............................7 Description of the Unit......................8 Functional description .......................9 Initial Operation ........................10 Preliminary Steps......................
  • Page 4 Table of Contents iii / iii SDI® 4 (G/54320/EN 0220 Rev01) Instruction Manual...
  • Page 5: Scope Of Delivery

    1 Scope of Delivery Scope of Delivery SDI® 4 Product (versions) SDI® 4m 54302  with two protective dust covers and transportation lock SDI® 4m (Leica Version) 54312  with two protective dust covers and transporting lock SDI® 4e 54300 ...
  • Page 6: Graphic Symbols Sdi® 4

    2 Graphic Symbols SDI® 4  If transport damage is discovered from the shipment, please file a complaint with the shipping company immediately.  Have the damage confirmed on the bill of lading so that an orderly handling of the complaint for damages is possible. Graphic Symbols SDI®...
  • Page 7: Safety Instructions

    3 Safety Instructions Safety Instructions Attention All safety-related instructions for use of the SDI® 4 are given in the Instruction Manual for the unit. It is therefore imperative that you read and understand the whole Instruction Manual before you use the SDI® 4. About this Manual ...
  • Page 8: Safety Instructions For Use

    Ensure that the SDI® 4 is used only by duly trained physicians and operating theatre personnel, who, due to their training or their knowledge and practical experience, can guarantee proper handling of the device. Only OCULUS or an authorized dealer is allowed to train the personnel. Transport and storage instructions Refer to the notes in sect.
  • Page 9 3 Safety Instructions Operation and Maintenance Information  Before initial operation: Let OCULUS or an authorized dealer train you in the operation of the SDI® 4.  Never operate a damaged SDI® 4. Inform the supplier of any damage. ...
  • Page 10 3 Safety Instructions EMC and Electrical Safety Information The SDI® 4c, a connected BIOM® 5 and a microscope constitute a Medical Electrical System (ME System) in accordance with DIN EN 60601-1. If you connect any more devices, they too become part of the ME System. ...
  • Page 11: Proper Use

    Only OCULUS or an authorized dealer is allowed to train the personnel. The SDI® 4 is designed for use in hospitals and clinics in conjunction with the appropriate operation microscopes.
  • Page 12: Description Of The Unit

    5 Description of the Unit Description of the Unit 1 Top dust cover 5 Bottom dust cover 2 Transporting lock for star knob 6 Indicator lamp 3 Star knob 7 Connection for a foot switch (optional) 4 SDI® 4e 8 Power supply connection Fig.
  • Page 13: Functional Description

    6 Functional description Functional description The SDI® 4 (Stereoscopic Diagonal Inverter) is used to reinvert images, e.g. that of a BIOM® 5 or a wide angle contact lens. SDI® 4e/4c On the SDI® 4e/4c, the image can be inverted by pressing an optional foot switch.
  • Page 14: Initial Operation

    Initial Operation Attention Risk of incorrect use due to improper training  Before initial operation: Let OCULUS or an authorized dealer train you in the operation of the SDI® 4. Preliminary Steps  Remove the plastic carrying case from its packing and dispose of the packing properly.
  • Page 15 This is guaranteed if the plug-in power supply unit provided by OCULUS is used. Make sure that the voltage supply in the mains socket is within the voltage range of the plug-in power supply unit.
  • Page 16 7 Initial Operation Connector SDI® 4c  Make sure that the plugs are plugged into the appropriate jacks or sockets.  Plug the plug for the foot switch into the 5-pole jack (3).  Plug the power supply plug (6V-15V) into the 3-pole jack (4). ...
  • Page 17: Operation

    8 Operation Operation Prior to each Use Before each use, check that  the unit is in technically perfect condition all connections and fasteners that can be loosened are properly  tightened and are in a safe condition all cables and plugs are free of damage ...
  • Page 18: Care And Maintenance

    Clean only the exterior surfaces of the unit with a damp cloth.  Do not clean the SDI® 4 with aggressive, chlorine-containing, abrasive or sharp cleaning agents. If the protective glasses are dirty, contact OCULUS Service or an authorized dealer. 14 / 24 Instruction Manual SDI® 4 (G/54320/EN 0220 Rev01)
  • Page 19: Sterilization Of The Rubber Cap

    10 Transport and Storage Sterilization of the Rubber Cap If you are using the rubber cap for the star knob, it must be sterilised first. The rubber cap for the star knob of the SDI® 4 can be autoclaved with steam (max.
  • Page 20 Clean the glass surface of the are dirty BIOM® 5 BIOM® 5 loupe If the SDI® 4 optics are dirty,  contact OCULUS Service or your authorized dealer. The SDI® 4 or its fixtures wobble  The locking screws are loose ...
  • Page 21: Guarantee And Service

    SDI® 4 is improperly maintained or handled, then any liability by OCULUS is voided. If the housing of the SDI® 4 is opened, any liability by OCULUS is  excluded.
  • Page 22: Manufacturer's And Service Addresses

    13 Disposal of Used Devices 12.2 Manufacturer’s and Service Addresses Supplemental information is available from our Service Department or from our authorized representatives. Germany: OCULUS Optikgeräte GmbH Münchholzhäuser Str. 29 35582 Wetzlar GERMANY Tel.: +49 641 2005-0 Fax: +49 641 2005-299 E-mail: support@oculussurgical.de...
  • Page 23: Order Information, Accessories And Replacement Parts

    14 Order Information, Accessories and Replacement Parts 14 Order Information, Accessories and Replacement Parts SDI® 4 models SDI® 4 models Order Number SDI® 4m 54302   SDI® 4m (Leica version) 54312 SDI® 4e 54300  SDI® 4e (Leica version) 54310 ...
  • Page 24: Technical Data

    15 Technical Data 15 Technical Data Fig. 15–1: Dimensions SDI® 4c Serviceable life expectancy 6 Years SDI® 4e and SDI® 4m dimensions Width 96 mm Depth 91.4 mm Height 38.9 mm Classification according to IEC 60601 - 1 Type of protection against electric shock Protection class 2 Housing protection class IP20...
  • Page 25 15 Technical Data Degree of protection if used in the presence of potentially explosive mixtures The unit is not suitable for use in potentially explosive atmospheres or in the vicinity of potentially explosive mixtures of anaesthetics or laughing gas and oxygen. Ambient, transport and storage conditions Ambient conditions Transport conditions...
  • Page 26 15 Technical Data Switching options: SDI® 4e Combination foot switch (electric, plugged into a 5-pole jack at the SDI® 4e) Star knob (manual) Switching options: SDI® 4m Star knob (manual) Sterilization and disinfection procedures Combination foot switch Wipe-down disinfection Rubber cap for star knob Steam autoclaving (max.
  • Page 27: Datasheet For Plug-In Power Supply Unit Gem12I 09-P1J (54905)

    15 Technical Data 15.1 Datasheet for plug-in power supply unit GEM12I 09-P1J (54905) G E M 1 2 12W AC-DC High Reliability Interchangeable Medical Adaptor s e r i e s G E M 1 2 12W AC-DC High Reliability Interchangeable Medical Adaptor s e r i e s Interchangeable AC plug Specifically for GEM12 ~40 PUSH...
  • Page 28 Manufacturer and Service Address Germany: USA: OCULUS Optikgeräte GmbH OCULUS Surgical, Inc. Münchholzhäuser Str. 29 562 NW Mercantile Place, Ste. 104 35582 Wetzlar Port St. Lucie, FL 34986, USA GERMANY Tel.: +1 772-236-2622 (outside of U.S.) Tel.: +49 641 2005-0 Tel.: +1 855-SDIBIOM (Toll free in U.S.)

This manual is also suitable for:

Sdi 4cSdi 4eSdi 4m543025431254300 ... Show all