Page 2
If you have any questions or would like additional information about your device, please do not hesitate to call us or to contact us by mail or fax. Our service team will gladly assist. OCULUS Optikgeräte GmbH Article number: G/54320/EN Revision: 03 Release: 06.07.2023...
Table of Contents Table of Contents Scope of Delivery....................1 Graphic Symbols SDI® 4 .................. 3 Safety Instructions....................4 About this Manual.................. 4 3.1.1 Used Graphic Symbols ............4 Safety Instructions for Use..............5 Proper Use......................8 Description of the Unit ..................9 Functional description ..................10 Initial Operation ....................11 Preliminary Steps ..................11...
1 Scope of Delivery Scope of Delivery Product (versions) SDI® 4m 54302 with two protective dust covers and transportation lock SDI® 4m (Leica Version) 54312 with two protective dust covers and transporting lock SDI® 4m (Kaps version) 54305 ...
Page 5
1 Scope of Delivery Accessories for SDI® 4 c and SDI® 4 e Optional: Plug-in power supply GEM12I 09-P1J 54905 Optional: extension cable for voltage power supply plug (approx. 5 m) 54898 Required accessories: Combination foot switch SDI® 4c 10009585 ...
3 Safety Instructions Safety Instructions Attention All safety-related instructions for use of the SDI® 4 are given in the Instruction Manual for the unit. It is therefore imperative that you read and understand the whole Instruction Manual before you use the SDI®...
operating theatre personnel, who, due to their training or their knowledge and practical experience, can guarantee proper handling of the device. Only OCULUS or an authorized dealer is allowed to train the personnel. Transport and storage instructions Refer to the notes in sec.
Page 9
3 Safety Instructions Operation and Maintenance Information Before initial operation: Let OCULUS or an authorized dealer train you in the operation of the SDI® 4. Never operate a damaged SDI® 4. Inform the supplier of any damage. Only operate the SDI® 4 with the original accessories supplied by ...
Page 10
3 Safety Instructions Attention Use of a Multiple Socket Extension Cord Risk of personal injury or material damage caused by unsafe multiple socket extension cord If you use a multiple socket extension cord to connect the SDI® 4 to the power supply, you must heed the following information: Use a multiple socket extension cord that complies with the ...
5 Description of the Unit Description of the Unit 1 Top dust cover 5 Bottom dust cover 2 Transporting lock for star knob 6 Indicator lamp 3 Star knob 7 Connection for a foot switch (optional) 4 SDI® 4e 8 Power supply connection Fig.
6 Functional description Functional description The SDI® 4 (Stereoscopic Diagonal Inverter) is used to reinvert images, e.g. that of a BIOM® 5 or a wide angle contact lens. SDI® 4e/4c/4eli/4cli On the SDI® 4e/4c/4eli/4cli, the image can be inverted by pressing an optional foot switch.
7 Initial Operation Initial Operation Attention Risk of incorrect use due to improper training Before initial operation: Let OCULUS or an authorized dealer train you in the operation of the SDI® 4. Preliminary Steps Remove the plastic carrying case from its packing and dispose of ...
Page 15
This is guaranteed if the plug-in power supply unit provided by OCULUS is used. Make sure that the voltage supply in the mains socket is within the voltage range of the plug-in power supply unit.
Page 16
7 Initial Operation 1 Set screw of the SDI® 4 4 Connection jack for the Power supply ® 2 Connection jack for the BIOM 5 Combination foot switch for SDI®/ BIOM® 3 Plug for the combination foot switch 6 Power adapter including interchangeable plug adapter for SDI®...
8 Operation Operation Prior to each Use Before each use, check that the unit is in technically perfect condition all connections and fasteners that can be loosened are properly tightened and are in a safe condition all cables and plugs are free of damage ...
Clean only the exterior surfaces of the unit with a damp cloth. Do not clean the SDI® 4 with aggressive, chlorine-containing, abrasive or sharp cleaning agents. If the protective glasses are dirty, contact OCULUS Service or an authorized dealer. 15 / 24 Instruction Manual SDI® 4 (G/54320/EN – Rev03)
10 Transport and Storage Conditioning of the rubber cap If you are using the rubber cap for the star knob, it must be sterilized first. The rubber cap for the star knob of the SDI® 4 can be autoclaved with steam (max.
the BIOM® 5 are dirty BIOM® 5 loupe If the SDI® 4 optics are dirty, contact OCULUS Service or your authorized dealer. The SDI® 4 or its fixtures wobble The locking screws are loose Tighten the locking screws ...
SDI® 4 is improperly maintained or handled, then any liability by OCULUS is voided. If the housing of the SDI® 4 is opened, any liability by OCULUS is excluded.
12 Guarantee and Service 12.2 Manufacturer’s and Service Addresses Supplemental information is available from our Service Department or from our authorized representatives. Germany: OCULUS Optikgeräte GmbH Münchholzhäuser Str. 29 35582 Wetzlar GERMANY Tel.: +49 641 2005-0 Fax: +49 641 2005-299 E-mail: support@oculussurgical.de...
13 Disposal of Used Devices Disposal of Used Devices Dispose of the SDI® 4 in compliance with the applicable legal requirements. Comply with the hygiene and disposal regulations of the hospital or clinic. SDI® 4, electric In accordance with Guideline 2012/19/EU of the European Parliament and also the law of the Federal Republic of Germany on the Commercialization, Recall and Environmentally Compatible Disposal of Electrical and Electronic Equipment, old electrical and electronic...
14 Technical Data Technical Data Fig. 14-1: Dimensions SDI® 4c/4cli Serviceable life expectancy 6 Years SDI® 4e/4eli and SDI® 4m - dimensions Width 96 mm Depth 91.4 mm Height 38.9 mm Classification according to IEC 60601 - 1 Type of protection against electric shock Protection class Degree of protection against electric shock Type B...
Page 25
14 Technical Data Ambient, transport and storage conditions Ambient conditions Transport conditions Storage conditions Temperature +10°C to +35°C -40°C to +95°C -10°C to +55°C Humidity 30% to 75% 10% to 95% 10% to 95% Air pressure 800 hPa to 1060 hPa 500 hPa to 1060 hPa 700 hPa to 1060 hPa These values also apply for all other SDI®...
Page 26
14 Technical Data Switching options: SDI® 4m Star knob (manual) Sterilization and disinfection procedures Combination foot switch Wipe-down disinfection Sterilisable cap for SDI® 4 Autoclaving, (max. 134°C (273° F)) Conditioning Instructions BIOM® 5 and Accessories The unit can be attached to the following microscopes: Zeiss ...
14 Technical Data 14.1 Data sheet for plug-in power supply unit GEM12I 09-P1J (54905) G E M 1 2 12W AC-DC High Reliability Interchangeable Medical Adaptor s e r i e s G E M 1 2 12W AC-DC High Reliability Interchangeable Medical Adaptor s e r i e s Interchangeable AC plug Specifically for GEM12 ~40 PUSH...