11 Troubleshooting
Fault
No function when the hand
switch / foot switch is pressed
(hand-held pushbutton for SDI® 4c
only optional)
The image is cropped
Unclear image
The SDI® 4 or its fixtures wobble
Fundus view is too narrow
Pronounced reflections on BIOM®-
loupe or contact lens
16 / 24
Fault list - SDI® 4
Possible cause
The hand switch / foot switch
is not connected to the
SDI® 4e/c
no connection of the SDI® 4e/
c to the power supply
Power failure or power socket
is not active
These are not active if the
sockets of the stand are being
used.
The SDI® 4 or other
components are incorrectly
mounted at an angle
The optics of the SDI® 4 are
not in operating position
(locking position)
The SDI® 4 or the optic of the
are dirty
BIOM® 5
The locking screws are loose
Distance between
ophthalmoscopic lens and eye
too large
(only when using
BIOM® 5)
Magnification of the
microscope system to high
The microscope light is on
Bright ceiling illumination
Instruction Manual SDI® 4 (G/54320/EN 0220 Rev01)
Remedy
Establish the connection to
the SDI® 4e/c
Establish the connection to
the power supply
Inform the in-house
electrician
Use the plug transformer
Switch the SDI® 4
mechanically
Activate the sockets in
accordance with the
instructions for the stand
Ask the microscope
manufacturer for assistance
Correct the assembly
Turn it on or off again with
the switch
Clean the glass surface of the
BIOM® 5 loupe
If the SDI® 4 optics are dirty,
contact OCULUS Service or
your authorized dealer.
Tighten the locking screws
Carefully reduce the distance
using the microscope fine
adjustment mechanism
Reduce magnification of the
microscope
Turn the light off, illuminate
only intraocular