Wall Mounting; Bench Mounting - Barnstead International MEGA-PURE A440697 Operation Manual And Parts List

6a water still
Table of Contents

Advertisement

Warning
Wall composition and
construction, as well as fastener
type, must be considered when
mounting this unit. The mounting
surface and fasteners selected
must be capable of supporting a
minimum of 110 lbs.; inadequate
support and/or fasteners may
result in damage to mounting
surface and/or equipment. If you
are unsure of mounting surface
composition, condition and
construction, or correct fasteners,
consult your building
maintenance group or contractor.
Caution
Replace front protective cover
before proceeding.

Wall mounting

The MEGA-PURE
holes for wall mounting. To support the weight of
this unit, use 1/4" x 1 1/4" lag bolts for mounting.
An optional wall mounting bracket is available for
the Automatic Collection System. Barnstead
demineralizers have slotted holes for wall
mounting. Demineralizer location is not critical,
though it should be close enough to the still to
make all necessary connections and should not
be mounted over electrical equipment to avoid
shock hazard (see warning above).

Bench mounting

No special consideration for a "still only" setup.
When used with the ACS, the MEGA-PURE
Water Still must be placed on a 12" platform to
allow for gravity feed of distilled water to the ACS.
Figure 10
®
6A Water Still has slotted
I
NSTALLATION
®
6A
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mega-pure a440518

Table of Contents