Tap Water/Demineralizer - Barnstead International MEGA-PURE A440697 Operation Manual And Parts List

6a water still
Table of Contents

Advertisement

Figure 8
Tap Water Demineralizer
Tap water/Demineralizer - Figure 8
(Must have the optional still adapter kit: catalog
no. 440376).
1. Thread the solenoid valve ("in" side
towards shut-off) onto the customer
supplied shut-off.
2. Install 1/4" O.D. tubing and the sup-
plied fitting to the "out" side of the
solenoid valve.
3. Cut the 1/4" O.D. tubing from the last
step at a convenient point and install
the tee from the still adapter kit.
4. Connect the 1/4" O.D. tubing from the
demineralizer to the tee.
5. Locate the boiler fill tube (Q). Connect
3/8" I.D. vinyl tubing from the deminer-
alizer output to the end of the boiler fill
tube. Route as shown and install clip
(T) to support.
6. Route the 1/2" I.D. vinyl tubing to an
atmospherically vented drain.
7. Plug the solenoid valve power cord
(not shown) into the receptacle in the
control box (O).
8. Plug the still adapter kit cable into the
receptacle located on the right side of
the demineralizer and the receptacle
in the control box (O).
P
D
LUMBING
ETAILS
13