Automatic Collection System; Plumbing Details; Tap Water Feed - Barnstead International MEGA-PURE A440697 Operation Manual And Parts List

6a water still
Table of Contents

Advertisement

Plumbing Details

Figure 7

Tap Water Feed

12

Automatic Collection System

Before connecting your MEGA-PURE
Still to the Barnstead Automatic Collection System
move the still to its final location.
Locate ACS tube (Z) and Thomas clamp (D) in the
parts box. Assemble to the tubing from the
collection system as shown in Figure 10.
Plug the input jack from the ACS into the ACS
receptacle on the bottom of the still control box (O).
See pages 12-14 for final plumbing details.
Plumbing Details
Tap water feed only - Figure 7
1. Thread the solenoid valve onto the cus-
tomer supplied shut-off, "in' side towards
shut-off.
2. Install 1/4" O.D. tubing and supplied fitting
to "out" side of solenoid valve.
3. Route 1/2" I.D. vinyl tubing to an atmo-
spherically vented drain.
4. Plug the solenoid valve power cord (not
shown) into the receptacle in the control
box (O).
®
6A Water