Pretreated Boiler Feed - Barnstead International MEGA-PURE A440697 Operation Manual And Parts List

6a water still
Table of Contents

Advertisement

A
SSEMBLY
Figure 4
Note
Refer to the Pretreated Boiler
Feed section if you will be
using a source of pretreated
water as boiler feed or to the
3 1/2 Gallon Storage Bottle
section if the optional 3 1/2
gallon storage bottle is being
installed. If neither applies,
move assembled still to its final
location. See pg. 14 for layout
recommendations, and pgs. 12-
14 for plumbing details.
10
7. Insert the immersion heaters into the top
openings of boiler (P) as shown in Figure
4.
8. Thread the heater plugs and cords through
the large hole above the boiler (P), then
back through the hole below the control
box (O). Plug the heater plugs into the
receptacles on the bottom of the control
box. The plug is a twist-lock type and
requires a 1/4 clockwise turn to lock in
place.
9. Route the water supply and the drain
tubing through the side of the cabinet.

Pretreated Boiler Feed

When a supply of distilled, deionized or reverse
osmosis water is to be used as boiler feed, revise
the still connections as follows:
1. Remove boiler fill tubes (V) and (Q) and
set aside.
2. Locate the 3/8" I.D. x 44" long vinyl tubing
and one barbed tee (1/2 x 1/2 x 3/8) in the
parts box.
3. Install one end of the vinyl tubing to the top
right tubulation on the condenser (B) and
route the other end through the cabinet
and down the back as shown in Figure 5.
4. Cut the 1/2" I.D. vinyl drain tubing and
install the barbed tee fitting. Connect 3/8"
I.D. vinyl tubing from the condenser to the
barbed tee . Push the vinyl tubing all the
way onto the tee to assure a leak free seal.