Lonza 4D-Nucleofector Instructions For Use

Inspection equipment for lv unit
Hide thumbs Also See for 4D-Nucleofector:

Advertisement

Quick Links

Inspection Equipment for 4D-Nucleofector™ LV Unit
This inspection equipment is a fixed resistor cartridge (560
(electrical performance) of the 4D-Nucleofector™ LV Unit prior to use.
AXP-1005
Inspection Equipment 4D-Nucleofector™ LV Unit
Test cartridge orientation: External part for handling (A)
and internal part for insertion into cartridge slot (B).
Test cartridge inserted into 4D-Nucleofector™ LV unit
Note: For a full IQOQ of the device, please refer to the respective IQOQ protocol.
› created: 2019-06-13
Ω resistor
Instructions for use:
1.
Switch on the 4D-Nucleofector™ System. If already
turned on go back to the home screen
2.
Select "LV Unit" and then "1 mL cartridge"
3.
Insert inspection equipment
4.
Select any solution code (needs to be done first to be
able to select a pulse code)
5.
Select the first pulse code setting indicated in the
table below (CA-100) and press OK
6.
Press START
7.
Observe the LEDs on the front of the inspection
equipment
8.
Check the result window on the screen (for expected
results see table below)
9.
Repeat steps 5-8 for remaining program codes (EN-
200, FF-100 and CA-137). Press the "Back" Button to
enter different program code
Pulse
code
CA-100
EN-200
FF-100
CA-137
IMPORTANT NOTE: Please do not use other pulses as the ones
indicated as this may damage the test equipment. If you want to use
another pulse for a certain reason, please contact Lonza to verify whether
that pulse might be suited.
› revised:
) that can be used to check the functionality
Expected Result
The six LEDs do not flash up.
No error message.
The six LEDs flash up twice each for more than
1 sec. No error message.
The six LEDs flash up twice each for less than 1
sec. No error message.
The six LEDs flash up twice each for less than 1
sec. No error message.
page: 1 of 2

Advertisement

loading

Summary of Contents for Lonza 4D-Nucleofector

  • Page 1 IMPORTANT NOTE: Please do not use other pulses as the ones indicated as this may damage the test equipment. If you want to use another pulse for a certain reason, please contact Lonza to verify whether that pulse might be suited.
  • Page 2 Quality Certification Inspection Equipment 4D-Nucleofector™ LV Unit Hereby Lonza Cologne, Germany confirms that this Inspection Equipment for the 4D-Nucleofector™ LV Unit has passed a quality control procedure comprising:  Optical check  Mechanical check  Electronic check  Safety Check Standards The Inspection Equipment 4D-Nucleofector™...