ENGLISH
10
Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
The "E" CODE is relative to errors of the operator, which can be corrected by same
operator.
The "A" CODE is relative to alarms, irregularities of the autoclave, after having carried out
the remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (CODE "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
To reset alarms and errors, maintain pressed the buttons here indicated until the
intervention of a
BEEP"
"
B-DOOR

10.1 List of error alarm messages

ERROR
E 200
Low water level
Low level water cleaned up with loadingfrom
E 250
demineralizer
E 300
Maximum level used water
E 401
Door open
E 451
Door unlocked
High temperature in chamber for vacuum test
E 461
(above 40° C)
A 250
Cycle begins without water
A 001
Interrupted cycle
A 011
Irregular pc display
A 101
Vacuum not reached in 10' (minutes)
A 111
Vacuum not maintened for first phase VT
A 121
Vaccum not maintened for second phase VT
Hatch opende during the cycle or problem
A 401
closing
A 501
Pressure not maintened
A 551
Irregular 1° temperature sensor
A 552
Irregular 2° temperature sensor
A 641
Overly high pressure
A 651
Overlay high temperature
A 661
Irregular temperature comparison
A 701
Working pressure not reached
A 751
Low temperature
A 761
Irregular pressor sensor
28

MESSAGES OF ERROR OR ALARMS

+
B-PUMP
CAUSE
operations.
+
B-SELECT
Fill tank
Fill tank
Empty tank
Colse door with attention
Open door and close again
Shut off and wait 10 minutes
Control demineralizer device
Reset and restart cycle
Shut off and restart
autoclave
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Check the closing system
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle
Reset and restart cycle