8.10 Interruption of sterilization cycle

A sterilization cycle can be voluntarily interrupted by pressing the button
B-START
for at least 2 seconds.
The autoclave will emit a beep, go into decompression and on the display will
appear the message ALLARM 001 (INTERRUPTED CYCLE).
To reset error, maintain pressed simultaneously the buttons below indicated, until the
intervention of an acoustic signal:
B-DOOR
+
B-PUMP
+
B-SELECT
ATTENTION:
You can't reset the allarm if the printer works
If an alarm message is displayed (CODE "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
operations.