3.4 Safety – Fig. 3.9
Warning. The Aquaprobe is provided with a safety
mechanism (see Fig. 3.9) which should be attached to its
securing collar as shown in Fig. 3.9B. This prevents rapid
outward movement by the probe if the nut 1 is released.
Note. To ensure maximum safety, the positioning
collar
be tightened in place using a 4mm hexagon
MUST
key.
A – Unsecured
B – Secured
Fig. 3.9 Safety Mechanism
3.5 Installing the AquaProbe – Figs. 3.10 and 3.11
AquaProbe suitable restraining equipment must be used
to prevent the probe being forced out under pressure.
Apply
PTFE Tape
See Text
3 MECHANICAL INSTALLATION...
Warning. When inserting or removing the
25mm (1in) Minimum
Clearance
Fig. 3.10 Insertion Bore Clearance
(hand tight only)
Insert Probe
Into Valve
Fig. 3.11 Installing the AquaProbe
Tighten
Tighten
Firmly
Remove
Cap
7