Z1 Clinician Programming Reference
The Z1™ has three settings that are clinician programmable:
1.
Device Pressure Setting
2.
Ramp Setting
3.
Z-Breathe™ Pressure Relief Algorithm Setting
Property of HDM. Confidential clinician information. Do not distribute.
ill. Z1 Interface Overview Diagram

Advertisement

loading

Summary of Contents for HDM Z1

  • Page 1 Thursday, November 21, 2013 | Z1 Clinician Programming Reference, v1 Z1 Clinician Programming Reference ill. Z1 Interface Overview Diagram The Z1™ has three settings that are clinician programmable: Device Pressure Setting Ramp Setting Z-Breathe™ Pressure Relief Algorithm Setting Property of HDM. Confidential clinician information. Do not distribute.
  • Page 2 Thursday, November 21, 2013 | Z1 Clinician Programming Reference, v1 Device Pressure Settings To set the treatment pressure: 1. Hold down the Start/Stop button while in Standby mode (display shows “0.0”).  The current treatment pressure number will flash on the display.  ...
  • Page 3 Ramp Settings To adjust the device ramp settings: 1. With the Z1 is in standby mode, press and hold the Up Arrow Button for approximately 3 seconds until the “00 MINUTES” value flashes . 2. While “00 MINUTES” value is flashing, repeatedly press either of the Up/Down buttons to increase or decrease the ramp time in 5 minute increments .
  • Page 4 Thursday, November 21, 2013 | Z1 Clinician Programming Reference, v1 Z-Breathe™ Algorithm Settings There are two Z-Breathe Algorithm settings: CPAP–1 Enables the Z-Breathe™ Algorithm at a gentle response rate which adjusts the airflow gradually during the inhalation-exhalation cycle. In this setting, the algorithm will initially adjust rapidly to breathing, but the adjustments will become gentle and gradual after a few inhalation-exhalation cycles.