All manuals and user guides at all-guides.com HYSTEROFLOW / HYSTEROBALANCE II HYSTEROSCOPIC FLUID MANAGEMENT SYSTEM QUICK REFERENCE GUIDE Option 1: Technical Assistance Option 2: Customer Service Option 3: Repair Option 4: Field Inventory...
Page 2
All manuals and user guides at all-guides.com W R N I N G This quick reference guide is only a summary of the HysteroFlow/HysteroBalance II Instructions for Use. For detailed operating instructions, be sure to follow the Instruction Manual that was included with your device when purchased.
Page 3
8. Clamp tubing and spike fluid irrigation bag(s). Open one clamp. HYSTEROFLOW QUICK REFERENCE GUIDE...
Page 4
Alarm Volume Menu Select a volume between 1 and 5. Bag Size Select a bag size between 0 and 5 liters. General Settings Menu To view and change Language, Sensor Test, Factory Defaults, Demo Mode, Brightness. HYSTEROFLOW QUICK REFERENCE GUIDE...
Page 5
3. Connect the short end of the Y tube to the patient drape. 4. Connect the long end of the Y tube to the “Patient” port of the first container. 5. Connect the last container tubing to suction. Outflow INSTRUMENT RECOGNITION (IR) HYSTEROFLOW QUICK REFERENCE GUIDE...
Page 6
Deficit Exceeded 10 Beeps In increments of 100 ml above the preset value Critical Overpressure 10 Beeps Actual Pressure > 200 mmHg above nominal pressure Rapid Deficit 10 Beeps Deficit change > 300 ml in 1 minute HYSTEROFLOW QUICK REFERENCE GUIDE...