HYS TEROFLOW / HY STEROBALANCE II HYS TEROS COPIC FLUID MANAGEMENT SYSTEM QUI CK RE FE RENCE GUI DE Option 1: Technical Assistance Option 2: Customer Service Option 3: Repair Option 4: Field Inventory...
Page 2
W R N I N G This quick reference guide is only a summary of the HysteroFlow/HysteroBalance II Instructions for Use. For detailed operating instructions, be sure to follow the Instruction Manual that was included with your device when purchased.
Page 3
C O N N E C T I N F L O W T U B I N G 1. Turn the unit on. 2. After self test, press Menu to check or change system settings. (Refer to page 4 for menu settings) 3.
Page 4
M E N U S E T T I N G S After each menu selection, press OK. Nominal Flow Menu Select a nominal flow value between 30 and 500 ml/min. Maximum Deficit Menu Select a deficit threshold value between 200 and 2500 ml.
Page 5
CONNECT OUTFLOW TUBING 1. Place containers on the bottom scale To Suction and toggle them with the connection tubes. 2. Connect the yellow Leur Lok on the outflow tubing to the outflow port of the instrument. 3. Connect the short end of the Y tube to the patient drape.
Page 6
B E G I N T H E P R O C E D U R E W R N I N G S Warning Message Alert Cause Change Fluid Bag 3 Beeps > 500 ml remaining Deficit Reached 3 Beeps Deficit exceeds the preset value Overpressure 3 Beeps...
Page 7
S Y S T E M C C E S S O R I E S Part # Description WA40634A Inflow and Outflow tubing set, Sterile, 5/pkg WA40635A Inflow tubing set, Sterile, 10/pkg 006775-901 Under Buttocks Drape/Fluid Collection Set, 5/pkg 006772-901 3L Canisters 36/pkg WA40634A...