Before starting any service work disconnect the cooling unit from power sup- ply! Never carry out any work at the refrigeration cycle, this work must be carried out at manufacturer site! COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 3 of 16 Tel. +41 31 744 9000, www.comet-xray.com...
Note: If the ambient temperature is below the adjusted anti freezing temperature the refrigeration system does not operate! Anti freezing, temperature maximum and bypass valve is factory adjusted according to specification. COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 5 of 16 Tel. +41 31 744 9000, www.comet-xray.com...
mount cover Operation on different mains voltages The installed transformer makes operation on either 208V 50/60Hz or 230V 50/60Hz possible. COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 8 of 16 Tel. +41 31 744 9000, www.comet-xray.com...
Scale down -: move contactor head against the direction of the oil flow Screw on fixing screw of contactor head COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 9 of 16 Tel. +41 31 744 9000, www.comet-xray.com...
Page 10
Note: Hysteresis of anti freezing T2 is adjusted at 4°C. If temperature drops below anti freezing temperature (+5 °C) refrigerating unit runs only if temperature exceeds +9 °C. COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 10 of 16 Tel. +41 31 744 9000, www.comet-xray.com...
Page 11
Release SET-key (P1 is shown on display) to run-mode (if there is no changing of parameters, the temperature controller returns automatically into run-mode after 45 seconds). COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 11 of 16 Tel. +41 31 744 9000, www.comet-xray.com...
10 minutes to deaerate cooling cir- cuit. Check filter for leakage Mount cover of the cooling unit COMET AG Herrengasse 10, CH-3175 Flamatt XRCA-3001-WA, Doc.-No. 50005554 Page 12 of 16 Tel. +41 31 744 9000, www.comet-xray.com...