Leibinger JET2neo Instruction Manual And Quick Start Manual

Leibinger JET2neo Instruction Manual And Quick Start Manual

Hide thumbs Also See for JET2neo:
Table of Contents

Advertisement

Leibinger V check
Instruction Manual and quick start guides for JET2neo
V. 1.2
05 / 2019

Advertisement

Table of Contents
loading

Summary of Contents for Leibinger JET2neo

  • Page 1 Leibinger V check Instruction Manual and quick start guides for JET2neo V. 1.2 05 / 2019...
  • Page 2 Page 2 V. 1.2...
  • Page 3  Activate the basic settings of the V-check in the printer  Adjust the V-check parameter for the selected printjob  Load the printjob Reference: chapter 5.1 Start and adjust the Leibinger printer 48 Step 3: Adding a new Inspection ...
  • Page 4: Table Of Contents

    Height adjust of the camera ..................... 27 4 Overview basic settings for printer and V-check ..............28 4.1 Basic settings for the Leibinger JET2neo printer ..................28 4.1.2 Activate the V-check in the respective print job ................30 4.2 Basic settings V-check ........................32 4.2.1 Overview about the respective menus and settings ..............
  • Page 5 5.2 Install the V-check and start the inspection ..................62 5.2.1 Adjustment of the discrete I/O ....................62 (Only required for initial startup or if the V-check don`t work correctly)..........62 5.2.2 Adjustments for the Startup Job ....................64 5.2.3 Configuring of the Match Sensor....................
  • Page 6: Introduction

    Introduction Publisher For questions regarding the operation and running of the Leibinger V-Check as well as in service case please contact the listed dealer address. Dealer address Paul Leibinger GmbH & Co. KG Marking&Coding Systems Daimlerstraße 14 D-78532 Tuttlingen Federal Republic of Germany...
  • Page 7 Consequently no guarantee can be assumed for the correctness of the content of this manual and no claims can be asserted against Paul Leibinger GmbH & Co. KG. We shall be grateful for information regarding possible printing errors as well as for suggestions for the further optimisation of the device manual.
  • Page 8: Structure Of The Manual

    Structure of the manual We are pleased that you have decided for a Leibinger V-check and welcome you as one of our customers. This manual shows you the fundamental advantages of the Leibinger V-Check system, such as for example the fully automatic working, low maintenance etc..
  • Page 9: Safety

    Safety Scope of risks The V-Check System has been built in accordance with state-of-the-art standards and recognized safety requirements and has been equipped with protective devices. Operational and safety checks of the installation were carried out before it left the factory. In case of improper handling or misuse, however, there are dangers for ...
  • Page 10: Safety Instructions

    Failure to observe this instruction can lead to material damage, the loss of guarantee and to injury! Indicates recommendations for use and other useful pieces of information. Indicates service activities! This work must only be carried out by trained personnel or by Leibinger service technicians! Page 10 V. 1.2...
  • Page 11: Intended Use

    Intended use The V-Check System was developed to record pictures from Leibinger CIJ printouts and evaluate them in environments with different brightness. The intended use of this device also includes the observance of all instructions in this manual. Using the installation for other purposes is considered contrary to its...
  • Page 12: Operation Staff

     Make sure that they have read and understood them. Dangers due to electric energy The electrical and electronic components of the V-Check System are under voltage. The device must only be opened by trained personnel or by Leibinger service technicians. Risk of temporary blinding There is a risk of being blinded temporarily.
  • Page 13: Safety Measures At The Place Of Installation

    Safety measures at the place of installation Operating environment! The installation is only allowed in a permissible operating environment. The device hast to be kept away from any inadmissible humidity, source of heat, flames and sparks. Attention : The complete device of the V-check including the V-check, the display and especially the focusing window are not consistent against any kinds of solvent!!! 2.10 Prior to assembly...
  • Page 14: Conformity

    2.11 Conformity In dependence of the original declaration of conformity Page 14 V. 1.2...
  • Page 15 V. 1.2 Page 15...
  • Page 16: Component Description And Installation

    Component description and Installation Description The V- check system is used in combination with a Leibinger printer. The system monitors and evaluates the printouts. The V-check system provides evaluation signals which are processed by the Leibinger printer. In the following chapter, the components of the V-check will be described.
  • Page 17: Technical Data

    Technical data Feature Description Screen size 3.5 zoll diagonal LCD aspect ratio Resolution 320 x 240 RGB 60 degrees left/right; Viewing angle 50 degrees up and 55 degrees down Display: Polycarbonat Sensor: dye cast zinc/PBT Material Bracket: Stylus: Delrin Display: 340 Gramm Weight Sensor:...
  • Page 18: Overview Components Of The Camera System

    Overview components of the camera system The V-check consists of the following components: 1. Touch display 2. Image sensor 3. Holder with set of screws for mounting the display 4. Cable set with stainless steel holder 5. Monitor cable 6. Reflection tube Dimensions (Display) All dimensions in millimeters.
  • Page 19: Dimensions (V-Check)

    Dimensions (V-check) All dimensions in millimeters. [Inches] unless otherwise noted Illustration 2: Dimensions Camera V. 1.2 Page 19...
  • Page 20: Assembly Drawing

    Assembly drawing Installation holder and display 2x Fillister screw 2x Mounting holes X4 Interface: Digital inputs X3 Interface: Digital outputs and Voltage supply Installed stainless steel holder Page 20 V. 1.2...
  • Page 21 Installation image sensor Installed sensor- variant 1 Information At variant 1 the printout is printed normally but it is depicted upside down in the display. The reason therefor is, that the picture can`t turned inside the sensor to depict it congruously. If you want to display the printout in the display also without upside down you must execute the following steps: ...
  • Page 22 Installed sensor- variant 2 Information At variant 2 the printout is depicted normally without upside down. Standard presentation Printout normally Pos. Description Quantity Fillister screw M4x6 Hex nut M6 Holder for image sensor Spacer bolt (long) Holder printhead Hexagon socket head cap screw M6 x 80 Hexagon socket head cap screw M6 x 60...
  • Page 23 Installation reflection tube on the camera (optional) Information The additional mounted reflector tube will focus the light of the built-in flashlight. This will improve luminance efficacy. In addition the light becomes more diffuse. The usage of the reflector tube is recommended when the luminous power of the built in flash is not sufficient or in case if the application requires a diffuse illumination.
  • Page 24: Image Sensor Component Description

    3.7.2 Image sensor component description The V-Check sensor is completely delivered with lens and integrated ring light. By the Remote Display the sensor is set up and monitoring inspections. If the integrated ring light is not used you need another light source. Please note Its possible to order a lense cap made of glass instead of polycarbonate.
  • Page 25: Switch On And Off Of The V-Check

    Information The V-check system has no separate on- / off button. The system is start up, by switch on the Leibinger printer and shut off by switch off the printer. Furthermore you have the following alternatives to restart the system during the...
  • Page 26: Led Indicators Of The Image Sensor

    LED indicators of the image sensor  In normal operation, the Power LED is steady green.  The Pass/Fail LED is green or red, depending on the triggered inspection.  The Ethernet I/O LED will be lit or off depending on connection status. Power LED Green: Power Red: Error...
  • Page 27: Sensor Assemblies

    The mounting bracket lets you easily position and adjust the sensor on your line. Information For optimal control results Leibinger suggest to adjust the distance between the lens and the printout surface at 100 mm!!!  When the distance between the lens and the printout surface is very high, the quality of the resolution is too bad.
  • Page 28: Overview Basic Settings For Printer And V-Check

    The following chapter gives a short overview which settings must be done on the printer, to activate the V-check. In chapter 5.1 Start and adjust the Leibinger printer these functions are explained. Activate the basic settings of the V-check system The V-check system will activate in the basic settings in the respective menu.
  • Page 29 Enable V-check by clicking on the checkmark <> in the checkbox <Enable V-Check> (1) and insert the delivered license key. Confirm and close the menu with the button <confirm > (2). 1 – Checkbox <Enable V-Check> 2 – Button <confirm> Call up the I/O settings to show the inputs and outputs Information When the V-check system is enabled the 3 inputs and 1 output for the V-check are...
  • Page 30: Activate The V-Check In The Respective Print Job

    4.1.2 Activate the V-check in the respective print job After the activation of the basic settings you must activate and set the V-check parameter in the desired print job: Main screen <Edit current Job> (1) Jobeditor <settings> (2) > <V-check Parameter> (3). 1 –...
  • Page 31 Overview about the vision parameters Information The following image gives you a first overview about the specific V- check parameter. In the chapter 5.1.3.2 Overview about the V-check Parameter these settings are described more detailed. 1 – Trigger Watch display 4 –...
  • Page 32: Basic Settings V-Check

    Information The following section gives you a short overview about the most important menus and their settings on the V-check, which are relevant in combination with the Leibinger printer. In chapter 5.2 Install the V-check these menus are detailed explained on a concrete example.
  • Page 33: Homescreen

    4.2.3 Homescreen The Home screen on V-Check Series sensor display is used to monitor inspections and to configure the sensor. Normally, the part being inspected is centered on the screen with the feature of interest bounded by the Region of Interest (ROI). green annotations indicate the object passes, and the annotations indicate a failure.
  • Page 34 Main menu with annotations Click the display mode icon to show the image with the annotations on. The green or red areas indicate sensors that pass or fail in the ROI. Main menu without Annotations Click the display mode icon to see the image without the annotations from the sensors. Inspection Statistics To access the Inspection Statistics, click the <Display mode>...
  • Page 35 4.2.3.2 Button <Help> By click on the button <Help> the respective information for the actual selected menu is displayed. In this example the respective information for the <image sensor> menu is displayed. Close the notice window by click on the button <OK>. 4.2.3.3 Button Manual Trigger The Manual Trigger icon is located on the lower-right of the sensor display on the Home screen and is used to manually capture a new image.
  • Page 36: Overview Menu V-Check

    4.2.4 Overview menu V-check The Main Menu has four sections: Inspection: to modify inspection settings Imager: to run the Auto Exposure routine and to make adjustments to functions like exposure, gain, and strobe System: to select the sensor Type and to manage the device Logs: to configure and view System and Inspection Logs Inspections menu...
  • Page 37 Information Green colored menus: Relevant for operating in combination with the Leibinger printer Hell colored Menüs: Not relevant for operating in combination with the Leibinger printer V. 1.2 Page 37...
  • Page 38 Mainmenu Logs menu System menu Live mode Sensor Mode Demo mode View LOG Inspection Logs Setup Save to USB Sensor Configuration Load from USB System Logs Default Inspection Logs View LOG Communication Logs Sensor Information System Logs Setup Communication Logs Unlock Sensor Lock without keyword Lock Device...
  • Page 39: Icon-List V-Check

    4.2.5 Icon-list V-check 4.2.5.1 Icons for operating Icon Description The Main Menu icon is displayed on the bottom-left corner of the sensor display on the Home screen. It provides access to sub-menus that are used to set up the sensor. The Inspection menu icon is located on the Main Menu, and provides access to parameters that need to be set for the current and all stored inspections.
  • Page 40: Display Icons

    Icons for operating Icon Description The Touch Calibration screen displays the Touch Calibration point at various locations on the screen. Every time the icon displays, the user taps the center of the icon to calibrate the screen. The Zoom Out icon is located on the right of the screen and is used to reduce magnification of the image being displayed.
  • Page 41: Communications Log Icons

    4.2.7 Communications Log Icons Icon Description Port opened. Port closed. Indicates that the command has been processed without errors. Indicates that the incoming entry is stalled (no new bytes), or end-of-frame delimiter was not received, or client is not reading data on ethernet. If the response frame contains an error or is dropped, the log entry icons for the request and the response frames will be colored red, and the displayed error count will increment by one.
  • Page 42: Check Default Settings On The V-Check

    If not, adapt them. Attention The original default settings must not adjust. Otherwise the V-check in combination with the Leibinger printer don`t work correct anymore. The default settings must be set as follows:  Main Menu > System > Discrete I/O (Menu for default settings)
  • Page 43: Sensor Types

    A reference pattern might include alphanumeric characters, logos, or any other shapes. During an inspection, the print outs of the Leibinger printer will checked. Additionally, if there is more than one of the identified pattern, the number expected can be set.
  • Page 44: 4.2.10 Important Menus For The Configuration Of The Image Sensor

    4.2.10 Important menus for the configuration of the image sensor Stored inspection / Adding a New Inspection 4.2.10.1 Go to: Main menu > Inspection > Stored Inspections and click <Add New>. This menu manages the inspections. You can create new inspections, add or delete inspections.
  • Page 45 Exposure time 4.2.10.4 Main menu > Imager > Exposure In this menu you can adjust the desired exposure time. Adjust the field of (FOV) on max. value 4.2.10.5 Main Menu > Imager > FOV In this menu you define the desired size of the field of view V.
  • Page 46: 4.2.11 Set The Parameters Of The Match Sensor

    4.2.11 Set the parameters of the Match sensor Main menu > Inspections > sensors The separate parameter which are relavant for the operation with the Leibinger printer will adjust in this menu. Therefore click on the button <Match1> (1) In the following these are: ...
  • Page 47 Rotationrange The <Rotation Range> sets the expected rotation of parts or labels during an inspection. The range of the adjustment is from 0 to 180 degrees. Pass count This setting is used to define the pass- or fail results of an inspection. ...
  • Page 48: Execute Of A Complete Inspection

    In the following chapter the complete procedure for setup the printer and the V-check is described based on an example. Start and adjust the Leibinger printer The following chapter describes exactly which settings you have to take on the printer to activate the V-check system.
  • Page 49 Enable V-check by clicking on the checkmark <> in the checkbox <Enable V-Check> (1) and insert the delivered license key. Confirm and close the menu with the button <confirm > (2). 1 – Checkbox <Enable V-Check> 2 – Button <confirm> Call up the I/O settings to show the inputs and outputs Information When the V-check system is enabled the 3 inputs and 1 output for the V-check are...
  • Page 50: Create, Configure And Load A New Print Job For Evaluation With The V-Check

    <Editor> window. button <create new job> (2) 1 – Selection menu <Job> 3 – Button <Enter> 2 – Option <Job edit Touch the softkey <A> (1) to enter the desired text. (For this example Leibinger V check) Page 50 V. 1.2...
  • Page 51 Insert the specific V-check parameter for the created print job After you have inserted the desired print job you must open the selection menu <Settings> (1) and activate the option <V-check Parameter> (2) to set the respective adjustments there. 1 – Selection menu <Settings> 2 –...
  • Page 52 5 – Tab <V-check 2 and V-check 3> Page 52 V. 1.2...
  • Page 53 5.1.3.2 Overview about the V-check Parameters 1 – Trigger Watch display 2 – Offset products display 3. – Trigger delay display 4 – Checkbox V-check error = Print stop 5 – V-check bad signals display 6 – V-check too fast signals display Adjust Trigger watch (1) Information At the trigger watch, the trigger signals were send...
  • Page 54 Adjustment of the trigger divider Information To use the V-check reliable also by fast production speeds you can reduce the issue of the triggers with the trigger divider. By default the divider is adjusted at 1 because every PrintGo signal should release a trigger. By request you can set the trigger divider individual.
  • Page 55 Adjust the Trigger delay (1) Triggerdelay Distance between print head and sensor-monitoring area: The trigger delay ensures, that the trigger released exactly at the time the printout is positioned within the Field Of View (FOV). This guaranteed a successful capture of picture recording.
  • Page 56 Schematic presentation of the trigger delay In the following exampIe the trigger delay is explained based on a graphic for a better understanding. Printhead V-check PrintGo Sensor Product 55 mm 55 mm Trigger delay: Product distance Distance between Print head and Sensor PrintGo delay: Distance between PrintGo Sensor und Print head...
  • Page 57 Schematic diagram For a better understanding, the following steps depicted the process flows, based on an example and illustrate how the V-check system is integrated inside the periphery of the printer. 1. Step: Product registration The product P1 is registered by the PrintGo sensor.
  • Page 58 Adjust the number of “bad signals” (2) Adjust the number of “bad” signals until a warning and error message occurs in the display. In addition, by activate the checkbox <V-check error> (1) the printer stops immediately when the first error level is reached.
  • Page 59 Adjustment of the print parameter Now, check the settings for the print parameter: Got to <Job settings> and open the tab <Print Parameter position> It is useful to adjust the pre adjustments for the print parameter as you see in the picture below.
  • Page 60: Storage The Print Job

    5.1.4 Storage the print job Save the job under the desired name and close the jobeditor with the button <> (2). 1 – Option <Save Job as> 2 – Button <Save> 5.1.5 Load job to print Now load the before created print job 1 –...
  • Page 61: Advice "V-Check Is Activated

    When V-check was activated as well as in the <Basic Settings> and in the <Job Settings>, the “V-check label” occurs in the main menu: Information Now all necessary settings for the Leibinger printer are done. Now adjust the required settings for the V-check system in chapter 5.2. V. 1.2...
  • Page 62: Install The V-Check And Start The Inspection

    Just in case of an error or when the V-check was rebooted it could be necessary to set the pulse width back to the default settings again. In the delivery condition, the Leibinger V-Check system is completely adjusted so that you don’t need to change the default settings.
  • Page 63  Check the following menu options. The engagements must be set as depicted: Input Polarity Input Pullup Output Type Output 1 Output 2 Output 3 V. 1.2 Page 63...
  • Page 64: Adjustments For The Startup Job

    5.2.2 Adjustments for the Startup Job 5.2.2.1 Add a new inspection Information At first you must add a new inspection  Go to: Main Menu > Inspection > Stored Inspections and click <Add New>.  Select the Matchsensor for the new inspection, and click <Next> ...
  • Page 65 inspection menu. V. 1.2 Page 65...
  • Page 66: Configuring Of The Match Sensor

    The following example shows step by step how to arrange the V-check system to a print job. At first you must print out the job with the LEIBINGER Logo, which you have created in chapter 5.1.3 and place it correctly under the image sensor, so that they catch it completely.
  • Page 67 Adjust the correct focus Main menu > Imager > Focus Attention  Don`t unscrew the screw complete. Loose the screw only 2 x turn, so that the clamp is loose enough to turn the focusing window.  It is normal that the focusing window is a little bit heavy to turn. ...
  • Page 68  Adjust the Exposure time either with the softkeys or via the slider If needed, go to Main Menu > Imager > Auto Exposure to run the Auto Exposure routine a second time or adjust Gain and Exposure manually: Information ...
  • Page 69: Set The Parameters Of The Match Sensor

    Adjust the field of view (FOV) on max. value Main Menu > Imager > FOV  Adjust the green frame by push and pull to adjust the size of the Field Of View (FOV)  Maximum FOV: 752×480 pixel or 376×240 pixel (fine/coarse resolution) ...
  • Page 70 Click Anywhere within the ROI to select it. When selected, the ROI is bolded with resize and rotational icons in the corners.  Resize the ROI so that it surrounds just the feature of interest. In the Demo example, the feature of interest is the LEIBINGER logo. Page 70 V. 1.2...
  • Page 71  Click the <Teach> icon to teach the sensor this good reference part.  Touch the softkey to save the settings and return to the previous menu. Information  When the teach process was successful no other message occurs. If the error message „pattern is invalid“ occurs you must increase the exposure ...
  • Page 72 Important tips for a successful teach process  Take care for a constant lighting  You should only teach meaningful patterns where a Letter and a number only once appear.  Avoid to teach similar letters and numbers. (For example don`t teach B and a 3 when they are contain in the same expression. ...
  • Page 73 Adjusted percentage quotation Button for increase the Buttons for decrease the percentage quotation percentage quotation Cursor for increase or decrease of the percentage quotation  Touch the softkey to save the settings and return to the previous menu Information When running a Match inspection wi th annotations enabled, the sensor will highlight in green any pattern matches that meet or exceed the value specified for Percent Match.
  • Page 74 Adjust the Rotation Range The <Rotation Range> sets the expected rotation of parts or labels during an inspection. For example, a value of 45 means that the part may rotate 45 degrees in either direction from the reference part and still pass. The range of the adjustment is from 0 to 180 degrees.
  • Page 75  Now adjust the <Rotation Range> to respect the expected rotation of a test part.  Therefore click on the black border of the screen to call up the cursor The rotation range with the respective settings is depicted as follows: ...
  • Page 76 Set the Pass Count parameter  Click on menu item <Pass Count> to open the respective menu  The <Minimum Pass Count> is the minimum number of parts, labels, or features expected to fall within the specified criteria.  The <Maximum Pass Count> is the maximum number expected to fall within the specified criteria. These settings are used to determine the pass or fail result of the inspection.
  • Page 77: Example: Load A New Created Inspection

     Touch the softkey to save the settings and return to the previous menu. Timeout: Main menu > Inspections > Sensors > Match > Timeout  Click on the menu item <Timeout> to open the respective menu The amount of time the inspection will hit on an image. If it times out before the inspection is complete, it will fail.
  • Page 78: Changing Running Inspections

    5.2.6 Changing Running Inspections The V-Check system offers the opportunity to change the inspection jobs during a running operation. This is necessary when the patterns changes during the print job. C hange during a running inspection:  From the Home screen, click the yellow button <Inspection>in the top center of the screen that displays the currently running inspection to display all the stored inspections.
  • Page 79: Multiple Sensors

    Multiple Sensors Some applications require more than one sensor. New sensors can be added after an inspection has been created. Input parameters for each sensor can be configured individually. The device holds up to 30 inspections. Inspections may hold multiple sensors. Basically you can add as much sensors as you like to an existing inspection.
  • Page 80  Click on the <yellow down-arrow> button to access sensor management functions. Sensor Management Remove sensors Softkey: Rename sensors <Sensor Setup> Change execution order Icon Function Move selected sensor up in the execution order Move selected sensor down in the execution order Edit name of selected sensor Delete selected sensor Configuration of the additional sensor...
  • Page 81 Representation of additional sensors in the main screen After the adding of a new sensor inside the main screen an extra frame is depicted. The sensor added will be displayed as an additional frame with a broken line. Choice box Information If the frames are overlapping clicking within one of the frames will open a selection box.
  • Page 82: Comprehensive Functions

    Comprehensive functions Information In the following chapter all further general functions of the V-Check system are described. Also those menu items which are normally stayed unused. Attention It is not allowed to do any adjustments here but for the parameters which are necessary for the V-check!!! Description of the menus There are 4 main menus:...
  • Page 83 Sensor Management Sensor Setup Remove sensors Rename Sensors Change Execution order V. 1.2 Page 83...
  • Page 84 Match Menu This chapter is already described in chapter 5.2.4 Set the parameters of the Match sensor. The following description is only for completeness. Main Menu > Inspection > Sensors > Match When the sensor is configured as a Match sensor, set the following values: ...
  • Page 85: Menu Motion (Only For Area- And Blemish Sensor Available)

    6.1.1.4 Pass Count Main Menu > Inspection > Sensors > Match > Pass Count The Minimum Pass Count is the minimum number of parts, labels, or features expected to fall within the specified criteria; the Maximum Pass Count is the maximum number expected to fall within the specified criteria.
  • Page 86  Click on <activate> in the below located mask to start the motion. The motion-mode is depicted as follows: After the option <Motion> was activated, those button is depicted in the inspection menu and can be dialed now. Mainmenu > Inspection > Motion The following submenus will open.
  • Page 87 V. 1.2 Page 87...
  • Page 88 6.1.2.1 Number of Edges Main Menu > Inspection > Motion > Number of Edges On the Number of Edges screen, use the radio buttons to select One Edge or Two Edges. If One Edge is selected, motion is tracked in one direction (by default, horizontally). If Two Edges is selected, motion can be tracked horizontally and vertically.
  • Page 89: Properties Menu

    6.1.3 Properties Menu Main Menu > Inspection > Properties The <Properties> menu is used to select a Sensor Type and specify an Inspection Name. Additionally, if the Sensortype is Match or Sort, an option to define a timeout for the inspection. 6.1.3.1 Inspection Name Main Menu >...
  • Page 90: Stored Inspections

    6.1.4 Stored Inspections Main Menu > Inspection > Stored Inspections <Stored Inspections> is used to manage stored inspections. Management of stored inspections includes adding, deleting, and specifying which inspection should be defined as the Startup inspection. From the Stored Inspections menu click: ...
  • Page 91 6.1.4.3 Startup Main Menu > Inspection > Stored Inspections > Startup The <Startup> button allows you to select the inspection to use as the startup inspection. The selected inspection will automatically start after power up. 6.1.4.4 Delete Inspections Main Menu > Inspection > Stored Inspections > Delete The <Delete>...
  • Page 92: Imager Menu

    Imager Menu Some parts of this chapter are already described in chapter 7.5. The following description is only for a better overview and for completeness. Main Menu > Imager The <Imager> menu icon is on the Main Menu, and lists parameters that affect the characteristics of the captured image.
  • Page 93: Gain

    6.2.3 Gain Main Menu > Imager > Gain <Gain> is an electronic boost to the image signal. Increasing Gain by using the keys or moving the slider to the right side increases image brightness without increasing exposure time. Note that Gain brightens both: The light pixels and dark pixels and may reduce the image quality. 6.2.4 Trigger Main Menu >...
  • Page 94: Focus

    6.2.5 Focus Main Menu > Imager > Focus The Focus Number displayed at the bottom of this screen is used to fine-tune image focus. Loosen the lock on the lens cover, turn the focus ring on the sensor until the Focus Number peaks (or the image appears sharp), then lock the focus ring.
  • Page 95 6.2.6.1 External Main Menu > Imager > Strobe > External The External Strobe is a 5V output that can be used for an external light. Setting options are:  Always ON,  Always OFF,  Exposure Based. If Exposure Based is selected, then the external light is on during the time the sensor is capturing an image. 6.2.6.2 Internal Main Menu >...
  • Page 96: Fov (Field Of View)

    6.2.7 FOV (Field of View) Main Menu > Imager > FOV The <Field of View> (FOV) is the area that the sensor can see at a given working distance. The working distance is the distance from the sensor's lens cover to the part being inspected. See also chapter Height adjust of the on page 27.
  • Page 97: Resolution

    6.2.8 Resolution System > Imager > Resolution Two resolution options are available on the V-Check Sensor: Fine and Coarse. Low resolution (Coarse) 376×240 pixels (max. FOV) High resolution (Fine) 752×480 pixels (max. FOV) Fine resolution has 4 times more pixels than coarse resolution. Inspection performance may be affected when using Fine resolution, depending on the application.
  • Page 98: System Menu

    System Menu Main Menu > System The <System menu> icon is on the Main Menu, and is used to manage the sensor. The System menu provides for selecting:  Sensor Mode  Sensor configuration  Sensor information  Lock Device ...
  • Page 99: System Configuration

    6.3.2 System Configuration Main Menu > System > Sensor Configuration The Configuration menu options are:  Save sensor Configuration to the USB flash drive  Load sensor Configuration from the USB flash drive  Reset the sensor Configuration to defaults 6.3.2.1 Save to USB Main Menu >...
  • Page 100: Sensor Information

    6.3.2.3 Reset to Defaults Main Menu > System > Configuration > Reset to Defaults Reset all sensor configurations to the factory defaults. Attention This operation will remove all existing inspections and replace them with factory default settings!!! 6.3.3 Sensor Information Main Menu >...
  • Page 101: Lock Device

    6.3.4 Lock Device Main Menu > System > Lock Device This option provides for locking the sensor to prevent accidental modification of settings. When locked, the sensor only provides access to pass/fail statistics, as well as the ability to view logs and to save them to a USB device.
  • Page 102 Page 102 V. 1.2...
  • Page 103 6.3.5.1 Ethernet I/O Main Menu > System > Communications > Ethernet I/O The sensor's Ethernet communications can be used to send data out the Ethernet port as part of an inspection, and remote devices can communicate with the sensor. The Ethernet I/O screen is where IP Address, Subnet Mask, and Gateway settings are configured. Use the expand arrow next to each field to display a software keypad to enter values for each field.
  • Page 104 6.3.5.2 Serial I/O Main Menu > System > Communications > Serial I/O Set Serial I/O settings for:  Baud Rate,  Data Bits,  Parity Control and  Stop Bits on this screen. Clicking Status displays recent bytes transmitted through this channel. Port Status Main Menu >...
  • Page 105 Connection Main Menu > System > Communications > Industrial Ethernet > Connection The V-check supports up to 4 communication channels. Inside these mask the following connection can activate:  Deactivate: All industrial Ethernet connections are deactivated and receive and send no data anymore. ...
  • Page 106 Custom Map Export Main Menu > System > Communications > Industrial Ethernet > Map > Custom > (Save icon) To export the Custom Map, Click to save a text listing of the map (Filename: iVuIEMap.csv) to an attached USB drive. Status Main Menu >...
  • Page 107 Delimiters Main Menu > System > Communications > Command Channel > Delimiters In the <Delimiters> screen, there are three delimiter options that you can set:  Field Delimiter: Which determines what is used to separate data that the sensor is sending out to a remote device. ...
  • Page 108 Data to Export Main Menu > System > Communications > Data Export > Data To Export The <Data to Export> screen is used to determine the information included in a data export. Data will output in the order displayed on the screen. Output Format Main Menu >...
  • Page 109 Data Export Advanced Main Menu > System > Communications > Data Export > Advanced During the Data and Image export operation, the sensor's output channels might become full. This can occur if the sensor is producing export data (frames) faster than the data can be exported from the device or faster than the client is reading the channel export data (due to bandwidth limitations).
  • Page 110 Connection Main Menu > Communications > Image Export > Connection The Connection screen is used to enable or disable the Image Export.  Deactivate: The channel for the image export is deactivate and sends no data.  Ethernet: The sensor transmitted image data via Ethernet as soon as receive a trigger signal Image Type Main Menu >...
  • Page 111: Discrete I/O

    Image Export Advanced Main Menu > System > Communications > Image Export > Advanced During the Data and Image Export operation, the sensor's output channels might become full. This can occur if the sensor is producing export data (frames) faster than the data can be exported from the device or faster than the client is reading the channel export data (due to bandwidth limitations).
  • Page 112 6.3.6.1 Input Polarity Main Menu > System > Discrete I/O > Input Polarity The V Check has two input signals: — Trigger and — Remote Teach. Both of these signals are edge sensitive. The operation of these signals is dependent on the Input Polarity setting. ...
  • Page 113 6.3.6.3 Output Type Main Menu > System > Discrete I/O > Output Type Select NPN to configure the sensor's outputs to sink current. Select PNP to configure the sensor's output for source current. Output 1, 2, and 3 Main Menu > System > Discrete I/O > Output (1) Output 1, 2, and 3 are setup separately to improve flexibility and simplicity.
  • Page 114 The default setting is Pass and Latched for Output 1. If Pulsed is selected, the default Pulse width is 10 ms. Information For Latched the signal is active until the results of an inspection cause a change in the signal output. The selection: Sensor Pass or Sensor fail...
  • Page 115: Display Settings

    On the <Select Sensor(s)> screen, the left column check box allows adding sensor position to be part of the logic that activates the output. One or more sensor positions can be added. The right column on the table shows the sensor name(s) and their respective position from the current inspection for reference. The bottom drop list has two settings: ...
  • Page 116 6.3.7.1 Fail Hold Time Main Menu > System > Display Settings > Fail Hold Time The Fail Hold Time determines how long a failed image is displayed on the LCD so that you can see what failed. The sensor will continue to process any triggers and the inspection will continue normally. This time delay is just for the screen.
  • Page 117: Reboot Sensor

    6.3.7.3 Touch Calibration Main Menu > System > Display Settings > Touch Calibration Touchscreen Calibration may be necessary if the software does not correctly respond when an icon on the screen is pressed. The calibration routine aligns the touch screen's coordinates to the display behind it. Be sure to follow the prompts on the screen when executing the Touchscreen Calibration function.
  • Page 118: Logs Menu

    Logs Menu Main Menu > Logs The Logs menu icon is on the Main Menu, and is used to set up, view, and save Inspection, Communication, and System Logs. The following settings can adjusted and stored in this menu: Inspection Logs System Logs Communication Logs 6.4.1...
  • Page 119 6.4.1.1 View Logs Main Menu > Logs > Inspection Logs > View Logs Inspection Logs appear as a strip of film. You can select a frame to view a specific image. Use the icon in the lower right of the screen to save the logs to the USB flash drive. Click the <Clear>...
  • Page 120: System Logs

    6.4.2 System Logs Main Menu > Logs > System Logs The <System Log> contains configuration change information, other notifications, and any errors or warnings that may be encountered. The list is sorted in descending order with respect to time. The 'Time' associated with each event consists of the Hour Count (lifetime hours of operation) and the Up Timer (time elapsed since last boot).
  • Page 121 6.4.3.1 Industrial Ethernet Log Main Menu > Logs > Communication Logs > Industrial Ethernet Log The <Industrial Ethernet> logs the most recent Industrial Ethernet activity. Setup Main Menu > Logs > Inspection Logs > Setup Enable or disenable the detailed data logging. 6.4.3.2 Command Channel Log Main Menu >...
  • Page 122 Information An hourglass displays if an operation takes a particularly long time to complete; for example, during a long trigger. Each log entry includes a log <detail button> (arrow icon on right side of log entry) to display a detail view of the log entry.
  • Page 123: Multiple Sensors

    Multiple Sensors Some applications require more than one sensor. New sensors can be added after an inspection has been created. Input parameters for each sensor can be configured individually. The device holds up to 30 inspections. Inspections may hold multiple sensors. Basically you can add as much sensors as you like to an existing inspection.
  • Page 124  Click on the <yellow down-arrow> button to access sensor management functions. Sensor Management Remove sensors Softkey: Rename sensors <Sensor Setup> Change execution order Icon Function Move selected sensor up in the execution order Move selected sensor down in the execution order Edit name of selected sensor Delete selected sensor Configuration of the additional sensor...
  • Page 125  Click within the Sensor Setup area (Black button) to display the input parameter menu for that sensor. (In this example it is the Match2).  After that push the softkey to take over the parameter settings and return to the previous menu. Representation of additional sensors in the main screen After the adding of a new sensor inside the main screen an extra frame is depicted.
  • Page 126 Choice box Information If the frames are overlapping clicking within one of the frames will open a selection box. (Selection Match 1 or Match2). Here you can also call up the sensor parameters and set the respective adjustments how it is described in chapter 5.2.4.
  • Page 127: V-Check Communication Summary Of Ethernet And Serial

    6.5.3 V-Check Communication Summary of Ethernet and Serial The V-Check communicates with other devices via Ethernet or a UART serial communications port (RS-232). In order to establish an Ethernet connection to the sensor, the external device must be configured with the correct IP address and TCP port to communicate.
  • Page 128: Troubleshootings And Warnings

    (a sensor reboot is not required). If the error occurs again, you can try rebooting the sensor to see if that fixes the problem. If the problem persists, contact Leibinger customer support.  Check if the V-check is activated The V-check don`t react ...
  • Page 129: Support

    Support This section provides general Leibinger resources and specific documentation for installers and operators of this V-Check Sensor. Product Support and service Leibinger provides the following resources for quickly setting up and operating the device: PDF Documentation The product documentation is available in a convenient printable format (PDF) on the provided CD.
  • Page 130: Maintenance

    Maintenance  Keep the operating area always clean in order to ensure a flawless operation of the V-check system.  The V-check system is free of maintenance.  It suffices to clean the housing from dust and dirt on a regular basis. You can use a neutral cleaning solution if required.
  • Page 131: Anhang

    Anhang 11.1 Index Add New .............. 86 Image sensor component description .... 24 Add Sensor ..........77, 118 Image Type ............105 Adjust the Field of View ........92 Imager menu ............36 Advanced ............112 Imager Menu ............88 assembly drawing ..........
  • Page 132 ROI Type ............67, 81 Startup..............87 Rotation .............. 84 Status ............98, 101 Rotation Range ..........72, 81 Stored inspection / Adding a New Inspection 44 Running Inspections ......... 76 Stored Inspections ..........86 Safety ..............9 Strobe ..............90 safety instructions ..........
  • Page 133: List Of Illustrations

    11.2 List of illustrations Illustration 1 Dimensions Display ........................18 Illustration 2: Dimensions Camera ........................19 Illustration 3: V-Check sensor ..........................24 Illustration 4: Connectors for the sensor ......................24 Illustration 5: Mounting holes ..........................24 Illustration 6: LED indicators ..........................26 Illustration 7: Video lens model ..........................