Auto False Echo Suppression

Enables a "learned" TVT curve to be used in place of the default TVT curve. Use this feature to
ignore false echoes on the echo profile. Set Range (Auto False Echo Suppression Distance)
first, then set Auto False Echo Suppression.
Range (Auto False Echo Suppression Distance)
Defines the endpoint of the Learned TVT distance.
1.
Rotate the instrument for best signal (lower false-echo amplitude).
2.
Go to Input > Detailed Setup > TVT setup > Distance.
3.
Determine the actual distance from the reference point (transducer face) to the material
surface.
4.
Subtract 0.5 m (20") from this distance, and enter the result.
Set Auto False Echo Suppression
1.
Open the menu Device – Auto False Echo Suppression and select the option to change it.
2.
Select Learn. The device will automatically revert to On (Use Learned TVT) after a few
seconds.
Display before Auto False Echo Suppression
false
echo
Display after Auto False Echo Suppression
false
echo
1.
This parameter cannot be reset to the factory default.
A5E32081626
SITRANS Probe LU (PROFIBUS PA) – QUICK START MANUAL Page EN-15
default TVT
true
curve
echo
Distance (meters)
TVT curve
material
(learned)
level
Distance (meters)