  • Page 4: Thm500 Series Instruments

    THM500 Series Instruments The Tektronix THM500 Series instruments combine the features of a complete digital multimeter (DMM) with the power of an oscilloscope, all in a single package. The DMM portion of these instruments measure true RMS AC and DC voltage, resistance, continuity with audible mode, and a diode test.
  • Page 5 THM500 Series Instruments THM500 series instruments will also store waveforms, settings, and displays. With these features, you can do the following: Recall settings Automate test procedures Save waveforms and screens in the field for later analysis Transfer waveforms and screens to a personal computer...
  • Page 6: Thm500 Series Specifications

    THM500 Series Specifications The characteristics listed in this section apply under the following conditions: H The instrument operates within the environmental conditions specified in Table 4 on page 8, unless otherwise noted. H The instrument warms up for at least 5 minutes.
  • Page 7 THM500 Series Specifications Table 2: DMM specifications (cont.) General Additional Readouts MIN: Minimum voltage or resistance MAX: Maximum voltage or resistance MAX–MIN: Difference between MAX and MIN HOLD: Value of the main reading when the HOLD button is pressed HOLD:...
  • Page 8 THM500 Series Specifications Table 2: DMM specifications (cont.) Diode test Range 0 to 2 V. Red input connector is positive. Continuity check Indication Audible tone and graphic of a closed switch displayed when resistance is <50 W (typical). Indicators optionally disabled.
  • Page 9 THM500 Series Specifications Table 3: Scope specifications (cont.) Vertical Mathematics Subtract: (CH 1 – CH 2) Add: (CH 1 + CH 2) Acquisition Modes Sample (Normal), Spike Detect, Roll, Run/Hold, Smooth, Dynamic DSP Acquisition Rate All modes except Dynamic DSP:...
  • Page 10 THM500 Series Specifications Table 3: Scope specifications (cont.) Waveform display Display Update Rate Dynamic Display Digital Signal Processing maps up to 750 waveforms/s on screen, simulating an analog-like display. Memories Setups Each setup memory stores the complete state of the instrument, including the multimeter state.
  • Page 11 THM500 Series Specifications Table 4: General specifications (cont.) Environmental Temperature Operating: 0_ to 50_ C (32_ to 122_ F) Storage: –20_ to +70_ C (–4_ to 158_ F) Humidity, Operating 0_ to 40_ C (32_ to 104_ F): Up to 90% relative humidity noncondensing (60% for 4 and 40 M measurements) 41_ to 50_ C (106_ to 122_ F): 60% relative humidity noncondensing.
  • Page 12: Thm500 Series Performance Verification

    THM500 Series Performance Verification This section contains procedures to verify that the THM500 Series instruments perform as warranted. Verify instrument performance whenever the accuracy or function of your instrument is in question. The performance verification procedures provide a valid confirmation of...
  • Page 13: Test Equipment

    THM500 Series Performance Verification Test Equipment The performance verification procedures use external traceable test equipment to directly check warranted characteristics. Alternative test equipment must meet or exceed the intended minimum requirements specified in Table 6. If you substitute equipment, you may need to modify the procedures.
  • Page 14: Functional Test

    Adapter or the THM5AC Power Adapter optional accessories to provide power. 2. Press the ON/OFF button to turn the THM500 series instrument on. 3. Press the MENU ON/OFF button to display the menus. NOTE. The THM575 can display the menus in more than one language. The language selections are located in the Utility menu.
  • Page 15: Dmm Verification

    The following check verifies DC voltage accuracy. WARNING. This test uses hazardous voltages. If you use standard banana plugs to apply test signals to the THM500 series instrument, do not touch the exposed conductors. 1. Set the THM500 series instrument to METER mode displaying V-DC. If the range display at the lower-right does not indicate AUTO, press the buttons simultaneously to set the RANGE to AUTO.
  • Page 16 The following check verifies AC voltage accuracy. WARNING. This test uses hazardous voltages. If you use standard banana plugs to apply test signals to the THM500 series instrument, do not touch the exposed conductors. 1. Set the THM500 series instrument to METER mode displaying V-AC. If the range display at the lower-right corner does not indicate AUTO, press the buttons simultaneously to set the RANGE to AUTO.
  • Page 17 Resistance Accuracy The following check verifies resistance accuracy. 1. Set the THM500 series instrument to METER mode displaying . Connect the THM500 series instrument to the resistance source as shown in Figure 3. cable BNC to dual banana cable Resistance...
  • Page 18: Oscilloscope Verification

    1. Set the THM500 series instrument to SCOPE mode. If necessary, press the SELECT button to point to CH 1. 2. Use a patch cord to short the THM500 series instrument COM and CH 1 inputs as shown in Figure 4.
  • Page 19 Input Leakage Trace Shift The following check verifies an acceptable input leakage trace shift. 1. Use the THM500 series instrument menus to set the sampling mode to SAMPLE (NORMAL for the THM571); then turn the menus off. 2. Use a patch cord to short the COM and CH 1 inputs as shown in Figure 5.
  • Page 20 200 mV/div. Use the buttons to set the time base to 1 s/div. 5. Use the THM500 series instrument menus to set CH 1 CPLG to DC; then turn the menus off. 6. Press the THM500 series instrument CURSORS/TRIGGER button until the readout shows TRIG in the lower-right corner;...
  • Page 21 Figure 7: Vertical accuracy test setup 2. Use the menus to set the THM500 series instrument CH 1 CPLG to DC; then turn the menus off. WARNING. This test uses hazardous voltages. If you use standard banana plugs to apply test signals to the THM500 series instrument, do not touch the exposed conductors.
  • Page 22 100V/div –300 V +300 V 552 V to 648 V 4. If your THM500 series instrument is equipped with two channels, repeat Step 3 (all tests) above on CH 2. 5. Disconnect the test setup. Analog Bandwidth The following check assesses the analog bandwidth.
  • Page 23 Set the horizontal scale of the THM500 series instrument to 10 s/div and the frequency of the leveled sine wave generator to 50 kHz. c. Adjust the leveled sine wave generator amplitude to show the waveform height and cursor readout listed in the second column.
  • Page 24 THM500 Series Performance Verification 2. Connect the THM500 series instrument CH 1 input to the COM input using a patch cord as shown in Figure 9. cable termination BNC to dual banana cable Leveled Sine CH 1 Wave generator THM5xx Figure 9: Crosstalk between DMM and scope test Setup 3.
  • Page 25 CH 1 vertical scale to 500 mV/div and the buttons to set the time base to 200 s/div. 4. Use the THM500 series instrument menus to set the sampling mode to SPIKE DETECT; then remove the menus from the display.
  • Page 26 3. Use the THM500 series instrument UTILITY menu to select CH 1 OFFSET. This automatically adjusts the channel offset. 4. Use a patch cord to short the THM500 series instrument COM and CH 1 inputs as shown in Figure 11.
  • Page 27 THM500 Series Performance Verification Bench Test Instruments and Handheld Oscilloscopes Basic Service...