Battery Exchange Procedure - NDD EasyOne-CS Field Service Manual

Table of Contents

Advertisement

2030434-006 Rev. E

13.4 Battery exchange procedure

Please make sure that your working environment is electrostatic safe to avoid damage to the used
electrostatic sensitive electronic components
Open the front cover by removing the two main screws.
Remove the printed circuit board (PCB)
Do not remove the sensor tube (two additional screws)
Remove the Lithium battery (see figures below)
Insert only 3V BR 1225 batteries which are approved by ndd Medizintechnik AG (e.g. Panasonic
3V BR 1225)
Make sure that the + Sign form holder and battery are matched.
Enter the new battery by preventing any short circuit (use only your fingers)
Insert the PCB back to the EasyOne-CS case.
Attach the front half of the case (make sure no cable will be squeezed)
False
EasyOne-CS Field Service Manual
Correct
Page 23

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for NDD EasyOne-CS

Table of Contents