10 Device Error Codes

Internal errors are reported using the following screen:
The number xx indicates the type of error that is detected. The following list shows all error numbers
and a description of the internal error.
No.
Description
1
internal software error (wrong argument)
2
internal software error (unsupported)
3
internal software error (timeout)
4
internal software error (data inconsistent)
5
internal software error (write error)
6
internal software error (read error)
7
internal software error (not found error)
13
internal software error (WD timeout)
14
the ultrasonic receive amplitude is too low (check spirette insertion)
15
the ultrasonic receive amplitude is too high (check spirette insertion)
16
Bootloader software inconsistent (CRC failure)
17
User software inconsistent (CRC failure). Update EasyOne-CS device software.
18
Calibration data inconsistent.
19
Failure of flash memory.
20
Flow sensor transit time range check error. Possible reasons: a) Blocked sound path, e.g. caused by
wrongly inserted spirette, b) almost empty batteries.
21
ndd ASIC serial transmission error (STX expected)
22
ndd ASIC serial transmission error (sequence error)
23
ndd ASIC serial transmission error (unknown character)
24
Real time clock RAM error
25
Configuration data inconsistent (lithium battery failure possible)
49
internal software error (overrun error)
50
internal software error (timeout error)
51
internal software error (storage error)
52
internal software error (record structure error)
53
internal software error (address error)
54
Serial data flash or serial communication data inconsistent (CRC failure)
55
internal software error. Writing to the internal Flash failed (flash command sequence error)
56
internal software error. Erasing the internal Flash failed (flash block erase error)
57
internal software error. Writing to the internal Flash failed (flash program error)
65
internal software error. Writing to the internal Flash failed (flash block program error)
66
internal software error. Accessing the internal Flash failed (flash busy)
67
First record of database reached. Go beyond this record failed (FIRST_RECORD)
68
Last record of database reached. Go beyond this record failed (LAST_RECORD)
69
No valid record in the database found (NO_VALID_RECORD)
70
internal software error (UNKNOWN_ERROR)
144
Sensor serial communication error
EasyOne-CS Field Service Manual
ERROR
Device Selftest
Failure #xx
Please contact
service
Page 15