Maintenance Instructions

Routine Testing

It is recommended to test the system weekly by simulating spasm movements
on the bed for 15 seconds (the default time unless changed by a user via
programming) & ensure the alarm call is sent to the pager.

How to Clean the Devices

It is also recommended to regularly clean both units by wiping with cotton wool
pads moistened (compressed until dripping stops) with a mild detergent (0.5%
washing up liquid) solution or by using an alcohol or baby wipe. Avoid getting
any liquid into the micro-USB connector.
mini
Changing Battery's in the Companion
Remove the base of the monitor as per figure 3 above OR figure 7 & 8 below.
Also please notice the polarity marked into the case [Fig. 5] when replacing the
2x AAA batteries.
After changing the battery's, the monitor MUST be turned on again by pressing
the recessed TEST button and waiting for the 60 second test mode to finish.
Figure 5.
Figure 6.