cial disinfectant solution (gigasept
lows:
1. Remove the connecting hose with angled hose
2. Operate the pedal until the disinfectant solution
3. Take the hose out of the disinfectant solution,
Important
The time of exposure to the disinfectant solution
must be adhered to. Unsufficient disinfection can
lead to danger of infection.
The two-part hose system with adapters, secretion
cap and secretion bottle must also be cleaned in the
disinfectant solution after every use. All parts must
be rinsed thoroughly with clear water and dried suf-
ficiently after every disinfection. The secretion bottle
can be autoclaved at up to 121°C/250°F.
connection from the secretion cap and suspend
it in a disinfectant solution.
emerges from the discharge valve.
hold the vacuum pump high and again operate
the pedal until all the disinfectant solution is
forced out of the pump. In case of serious con-
tamination, we recommend a precleaning with
clear water. Procedure is the same as for rinsing
with disinfectant solution. The pump does not
need to be dismantled for this cleaning opera-
tion.
®
FF (new)) as fol-
Hygienic preparation EN
13