5. Hygienic preparation
12
EN
Hygienic preparation
This product contains disposable items.
items are intended to be used only once. So use
these items only once and do
Reprocessing disposable items may impair the func-
tionality and safety of the product and lead to un-
foreseeable reactions as a result of ageing,
embrittlement, wear, thermal load, the effects of
chemical processes, etc.
After every aspiration process, hygienic preparation
of the MANUVAC aspiration pump and the accesso-
ries used is essential. We recommend immediate
cleaning with clear water (on the spot if possible).
Wiping with disinfectant is sufficient for the external
surfaces.
Observe the instructions regarding use of disinfect-
ant. For immersion disinfection we recommend
®
gigasept
FF (new) and terralin
disinfection. You are recommended to wear suitable
gloves (e.g. household or disposable gloves) during
disinfection procedures.
For further information on hygienic preparation and
a list of all cleaning agents and disinfectants which
can be used, please see our Internet brochure at
www.weinmann-emergency.com.
Make sure you perform a functional check after
every hygienic preparation (see "6. Functional
check" on page 14).
If, despite the overflow protection device, secretions,
blood, mucus or the like should pass into the interior
of the pump (the inside of the connecting hose will
show signs of contamination), it is necessary to clean
the interior – the pump chamber – with a commer-
Disposable
reprocess them.
not
®
protect for wipe