1.3. Ensure safety

► Please read the user manual completely before the first use.
► Only use the analyzer for the intended use and within the parameters
specified in the technical data.
► Do not use any violence.
► Avoid falls.
► Do not put the analyzer into use, if the housing, power supply unit or sup-
ply leads are damaged.
► Do not store the analyzer together with solvents. Do not use
desiccants.
► Only carry out maintenance and service work on this analyzer as de-
scribed in the User Manual. Observe the prescribed action steps.
► Operate the analyzer only in closed, dry rooms and protect it from rain
and moisture.
► When operating the analyzer on mains power, operate it only with the
mains adapter supplied.
► Do not use the metal tube of the gas sampling probe or other metallic
parts / accessories as electrical conductors.
► The analyzer must not be located in the immediate vicinity of open fire or
great heat.
► The specified temperature range of the gas sampling probe must not be
exceeded, otherwise the probe tube and temperature sensor will be de-
stroyed.

1.4. Important general information EN50379 and VDI4206

This analyzer is not designed to be used for continuous measurements.
Before using the analyzer verify the condition of the various parts of the ana-
lyzer, such as the probe, the ambient air conditions, the condensate separa-
tor, star filter and the connectors for damage and/or blockages.
When starting up the analyzer it will take between 1 – 3 minutes to set to zero
depending on the condition of the sensors and of ambient.
The minimum zeroing time of the analyzer to achieve correct measurement
values can be expected by 1.5 minutes.
Exposure to acids; aggressive gases such as sulphur; vapours such as thin-
ners, gasoline, alcohol and paint, etc. can damage, reduce the life of, or de-
stroy the sensors.
The life of the sensors depends on how they are used, maintained and treated.
Typical average life expectations are: O
and NO - 3 years.
The use of the analyzer for regulatory purposes is subject to special regula-
tions (for example a periodical examination of the analyzer). Please obtain the
appropriate regulations from your local responsible authority.
MRU GmbH, D-74172 Neckarsulm
USER MANUAL OPTIMA 7 Biogas
- 3 years; CO - 2 - 3 years, H
2
S 2 years
2
7 / 72