1. About this document This is the Technical Reference Manual for SIRIUSi-XHS® device, version V 20-1. SIRIUSi-XHS® is a high performance, high speed line of SIRIUS® real time data acquisition hardware ...
TECHNICAL REFERENCE MANUAL 1.2. Online versions 1.2.1. Online SIRIUSi-XHS® technical reference manual The most recent version of this manual can be downloaded from our homepage: https://download.dewesoft.com/list/manuals-brochures/hardware-manuals In the H ardware Manuals section click the download link for the S IRIUSi-XHS® technical reference ...
2. Getting started This chapter will help you to install the software, connect your SIRIUSi-XHS® device to the PC via Ethernet or via USB, and will show you how to configure DewesoftX® software. To follow these steps, you need the following items: ...
TECHNICAL REFERENCE MANUAL 2.2. Connecting SIRIUSi-XHS® In this chapter, you can see the basic instructions for connecting SIRIUSi-XHS® devices over 3.0 USB and RJ-45 ethernet connection. Advanced connections are described in the following chapters. 2.2.1. Connection over 3.0. USB ...
2.2.2.2. Ethernet configuration on the PC As SIRIUSi-XHS® is a network device with static IP, you need to adjust the Network Card settings when connecting it via Ethernet. The ethernet port on your SBOX, data logger or PC should be configured in a ...
IP replace IP with IP that is set on your device. The result should be something like this: I mage5: ping IP in Command Prompt SIRIUSi-XHS® - V20-1 9/47 ...
Settings item in the pop-up to open the DewesoftX® settings dialogue. I mage6: DewesoftX® settings In case the XHS device is connected via USB, the device will automatically be listed under the Dewesoft ...
Page 11
Real-time controller option. If two Sirius-XHS devices are used, two different RT devices need to be added to the device list. SIRIUSi-XHS® - V20-1 11/47 ...
Page 12
Support of RT-devices is still an Experimental feature and it needs to be enabled under advanced -> Experimental settings. SIRIUSi-XHS® - V20-1 12/47 ...
Page 13
I mage9: Adding the Real-time controller under Experimental After that you will see a new Sirius-XHS or Dewesoft RT connection in the device tree. I mage10: RT device Connection ...
Page 14
DEVICE TECHNICAL REFERENCE MANUAL If the text by the device is disconnected, you should click on the device (Dewesoft RT Connection) , set ...
Page 15
selecting time source. When you have only one device you can select the Dewesoft DAQ devices and ...
DEVICE TECHNICAL REFERENCE MANUAL 2.2.4. Channel Setup SIRIUSi-XHS® When Dewesoft has started up, you will be in the Measure mode and see the Setup files list or you can ...
the measurement setup. 2.3.1. Help - Manual Note that this document is just a quick start guide. For detailed information about Dewesoft consult the ...
D ewesoft upgrade package (.dxu file) from the Dewesoft downloads page under ● the section Drivers. Copy the file into the Firmwares folder of your Dewesoft installation (e.g. ● D:\DewesoftX\System\X3\Firmwares). Connect the Dewesoft instrument to the PC and run Dewesoft X3. ...
DEVICE TECHNICAL REFERENCE MANUAL 2.6. Licensing SIRIUS-XHS or any other Dewesoft device already comes with an embedded Dewesoft license. You can ❶ ...
TECHNICAL REFERENCE MANUAL 2.7. Troubleshooting If your SIRIUSi-XHS® device is not found by DewesoftX® : If you did not restart Windows after the software installation, restart now ● Make sure that you have started DewesoftX® version Release Candidate 2020.2, Development ...
SIRIUS XHS, due to the nature of the chosen filter, both modes are perfectly aligned and synchronized. As on any Dewesoft devices, data can be synchronized to other data sources, such as vehicle bus interface, GPS, IMU, video, and others. ...
3.1.5. Power analyzer Dewesoft software will automatically calculate and store not only raw data from the voltage and current transducers but also all the power parameters: P, Q, S, D, Cos φ, power factor, P, Q, cos φ for each ...
Max. Output Current ±24 mA (±50 mA for 1 sec) Max. Sync-cable length 100 m (Master/Slave), 200 m (IRIG) 3.3. Single slice A single SIRIUSi-XHS® slice can have up to 8 measurement modules, currently HV, LV and ACC modules are supported. ...
TECHNICAL REFERENCE MANUAL I mage44: S IRIUS-XHS 8xACC 3.4. Single slice: Rear side The SIRIUSi-XHS® chassis has following connectors at the rear side: I mage45: S IRIUSi-XHS® Rear side Name ...
Power out connector: pin-out (2-pin LEMO female) Power Out connector (on the device): EXG.1B.302.CLL Mating connector (for the cable): FGG.1B.302.CLAD52Z The Power Out power plug can be used to daisy-chain several devices together. Daisy chain cable: L1B2m-L1B2f SIRIUSi-XHS® - V20-1 30/47 ...
HV modules. 3.5.5. Ethernet Connector Connector used on the device is a standard ethernet connector (RJ45). Standard ethernet cable with standard connector can be used to connect SIRIUS-XHS with a PC. SIRIUSi-XHS® - V20-1 32/47 ...
The Illustrations to the right show a USB-C and Standard size USB sockets. I mage52 :USB-C connector on the device and USB connector on the PC, SBOX SIRIUSi-XHS® - V20-1 33/47 ...
DEVICE TECHNICAL REFERENCE MANUAL 4. Module overview Currently HV, LV and ACC amplifierts are supported on the SIRIUSi-XHS® device. XHS-HV XHS-LV XHS-ACC Connectors BANANA DB9 BNC Counter version - - - Channels per slice ...
HV connector: pin-out (Banana plug) Warning It is mandatory to connect a ground cable to the GND connector of the SIRIUS® when you are working with high voltages. 4.2.2. HV: Voltage SIRIUSi-XHS® - V20-1 37/47 ...
DSUB 9, BNC (others on request) TEDS support Standard + DSI® adapters 1) In- must be within ±10V referred to GND(iso); for Ranges ≥ 10 V the DC value of In- is not rejected SIRIUSi-XHS® - V20-1 38/47 ...
DC or AC (1 Hz, 0.1 Hz) Input Impedance 1 MΩ Overvoltage Protection 50 V continuous; 200 V peak (10 msec) I E PE mode Excitation 4 or 8 mA Compliance voltage 25 V Output Impedance >100 kΩ SIRIUSi-XHS® - V20-1 40/47 ...
Dewesoft d.o.o. assumes no liability for the customer’s failure to comply with these requirements. All accessories shown in this document are available as an option and will not be shipped as standard ...
Page 43
Dewesoft sales and service office for service and repair to ensure that safety features are ...
Page 44
and warnings in this manual. SIRIUSi-XHS® - V20-1 44/47 ...
Dewesoft system can be requested. We retain them for at least one year, after system delivery. 6.3. Support Dewesoft has a team of people ready to assist you if you have any questions or any technical difficulties ...
Need help?
Do you have a question about the SIRIUSi-XHS and is the answer not in the manual?
Questions and answers