ECO Sensors UV-100 Instructions For Use Manual

ECO Sensors UV-100 Instructions For Use Manual

Ozone analyzer
Hide thumbs Also See for UV-100:

Advertisement

Quick Links

ECO SENSORS, INC.
OZONE ANALYZER
Model UV-100
INSTRUCTIONS FOR USE
GENERAL
The model UV-100 is a UV absorption analyzer
designed specifically for ozone. It uses a mercury
lamp filtered for absorption at 254 nm as is typical of
other commercial instruments. The Eco Sensors, Inc.,
design is very compact and uses the latest digital
microprocessor technology. The UV-100 covers three
ranges: 0-1, 0-10 and 0-999 PPM. Its outputs are a
LCD digital display, 0-2.5 VDC, 4-20 mA, and a time
and date stamped digital data stream for external
readout and data logging. The data output are via
serial port and USB. There is also an internal data
logger which can store and forward approximately 36
hours of data. The theory of operation is described in
Appendix D.
Panel LEDs indicate "on", low lamp output, low flow rate, and the ozone concentration at
the sampling point exceeds 0.1 PPM.
The instrument is conditionally warranted for one year. Read the warranty statement at
the end of this manual.
OPERATION
Plug the AC adapter into AC power and plug the adapter's output plug into the power
jack on the instrument. Find the power switch on the rear panel and turn the instrument
on. The green LED "on" indicator and the LCD display backlighting should illuminate.
Allow at least 1/2 hour warm-up. More than 1 hour is preferred if time permits. Initial
PPM readings may start out high and possibly drop to zero for a few minutes. Then the
readings will tend towards the actual value as the systems settles into equilibrium.
The compact, high-speed sample pump in the instrument makes a buzzing sound.
This is normal.
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UV-100 and is the answer not in the manual?

Questions and answers

Summary of Contents for ECO Sensors UV-100

  • Page 1 Model UV-100 INSTRUCTIONS FOR USE GENERAL The model UV-100 is a UV absorption analyzer designed specifically for ozone. It uses a mercury lamp filtered for absorption at 254 nm as is typical of other commercial instruments. The Eco Sensors, Inc., design is very compact and uses the latest digital microprocessor technology.
  • Page 2 Connect your sample feed tubing to the instrument's 1/4" input compression fitting or via a small optional sample line filter. The Tygon feed tubing is Teflon lined polyethylene so that the ozone is flowing through Teflon, but the tubing is more flexible than pure Teflon. The range is selected automatically by the instrument unless it is set to a preferred range by the Function control on the front panel.
  • Page 3 MENU SECTIONS Upon start up, the UV-100 will display software version number then date and time. Pressing the function knob will open the main menu. Scale (SCL) Menu: Pressing the function knob again enters the scale menu.
  • Page 4 The UV-100 will now store all readings at 10 second intervals until internal memory is filled. The data is retained even if the unit is powered down or unplugged so the UV-100 can be used remotely and returned to a computer for download.
  • Page 5 CAUTION: The relay is rated at 240V and 5A. For controlling large motors and other heavy electrical loads a power relay will need to be cascaded off the UV-100’s internal relay. Date and Time Adjust (D/T):...
  • Page 6: Panel Led Indicators

    Power: UV-100 is powered and operating. TRANSMITTING DATA TO COMPUTER Data from the UV-100 is easily transmitted to the PC using the Microsoft Windows® operating system. The program used for this is HyperTerminal found in Windows XP and previous versions.
  • Page 7: Service And Maintenance

    Data output via RS-232 Connector: This output is ozone concentration in PPM, temperature, pressure, and time and date sent in ASCI via the RS-232 connector on the rear panel of the UV-100. This can be directly connected to the serial COM port of many computers.
  • Page 8: Specifications

    If a defect develops during the warranty period, Eco Sensors at its election will repair the instrument or replace it with a new or reconditioned model of equivalent quality. In the event of replacement with a new or reconditioned instrument, the replacement unit will continue the warranty of the original unit.
  • Page 9 RECORD YOUR SERIAL NUMBER HERE__________________________________ KEEP THIS MANUAL AND WARRANTY FOR YOUR RECORDS. Eco Sensors is a registered trademark of Eco Sensors, Inc. © Eco Sensors, Inc. 2004. Revision 08/11 For brochures, application and tech notes, and other useful information, visit our extensive website at www.ecosensors.com.
  • Page 10 2. You will be asked for a CONNECTION DESCRIPTION. Click OK. 3. Choose which port you will use to connect to the computer. The default serial COM port on most PC computers is COM1. Click OK. 4. Enter PORT SETTINGS as above: Bits per second – 9600 data rate, Data bits - 8,...
  • Page 11 • USB to UART Driver Disk 1. Install USB to UART Driver Disk in the computer’s CD ROM drive. 2. With the UV-100 off, attach USB cable from UV-100 to a USB port on the computer. 3. Turn on UV-100.
  • Page 12 7. Run the Installation wizard. 8. Finish the driver installation.
  • Page 13 PINOUT OF DB9 CONNECTOR APPENDIX D APPLICATION EXAMPLES Here are 3 widely used examples installed by many users of the UV-100. These are shown for reference only; for more details contact Eco Sensors Tech Support. REMEDIATION EXAMPLE Ozone’s cleaning and disinfecting properties are well known. Its ability to break down bacteria and a wide range of contaminants with almost no residue is just beginning to be used in real-world applications.
  • Page 14 UV-100. Several methods can used to shield the UV-100 from theses extremes. The UV-100 is kept isolated by being outside the cold space and by using a long length of inlet tubing, the majority of which coiled outside the cold space. This allows for the inlet sample to be warmed to an acceptable level before measurement.
  • Page 15: Theory Of Operation

    USB cable. Set input type for channel 1 to Voltage, scale to 0 VDC = 0 PPM and 2.5 VDC = to the scale value in the UV-100. For further details about HOBOware setup contact technical support.
  • Page 16 Where l is the path length (6.2 cm) and σ is the absorption cross section for ozone at 254 nm (1.15 x 10-17 cm2 molecule-1 or 308 atm-1 cm-1), which is known with an accuracy of approximately 1%. The Eco Sensors instrument uses the same absorption cross section (extinction coefficient) as used in other commercial instruments.
  • Page 17 OPTIONS Available for UV-100 Pelican case PN 20-0598 $ 180.00 Hi Efficiency Sample Line Filter and Housing PN FLTRS(LARGE) $ 126.00 Hi Efficiency Sample Line Filter Elements 10pcs PN FILTERPAK(LARG) $ 63.00 Heavy Duty External Pump(replaces internal pump) PN EXT PUMP(HD) $ 440.00 240 VAC 50Hz Adapter $ 75.00...