Technical Data; Smart Radon Sensor - SARAD Smart Radon Sensor Manual

Radon monitor for building automation
Table of Contents

Advertisement

SARAD GmbH

Technical data

Smart Radon Sensor

Radon measurement
Operational principal
Sampling method
Sensitivity
Range
Uncertainty (1)
Ambient
Humidity
Range
Accuracy
Temperature
Range
Accuracy
1
Pressure
)
Range
Accuracy
1
CO
Sensor
)
2
Principle of operation
Range
Accuracy
Response time
Remark
1) Available as an option
Manual_Smart-Radon-Sensor_EN_12-03-2020.docx
Lucas cell with gross alpha counting
Diffusion
> 2,5cpm/(kBq/m³)
1 ... 1MBq/m³
1 hour @ 300Bq/m³
1 day @ 300Bq/m³
1 day @ 50Bq/m³
-10°C ... 50°C, 0%rF ... 100%rF non condensing
0%rH ... 100%rH
< 4,5%rH (3% typ.) for 20%rH ... 80%rH
-40°C ... 120°C
< 0,4°C (0,3°C typ.) for 5°C ... 60°C
760mbar ... 1200mbar
< 0,5% FSO
Non dispersive infrared (NDIR)
400ppm ... 5000ppm
< 5% ± 50ppm
10min
Automatic calibration with respect to outdoor CO2 level
20
15%
3%
8%

Advertisement