Restarting the
measuring operation
Actions if the filter plate
has dried out
ba76171e02
04/2019
4
Clean the filtration unit.
5
Dismount the filter module from the guide slide.
6
Unscrew the Allen screws of the frame (6 hexagon socket
screws).
7
Remove the used filter plate with the adapter.
8
Clean the frame parts.
9
Remount the new filter plate in the frame (see section
3 M
/
OUNTING
DISMANTLING OF THE FILTER MODULE
4 I
NSTALLING THE FILTER MODULE
10
Submerse the filtration unit in the basin or channel.
11
Switch on the filtration pump with the Start key.
Carry out the general steps to restore the readiness to measure of the
analyzer (see operating manual of your analyzer, chapter Mainte-
nance).
5.5

Storing a used filter plate

Proceed as follows when you want to store the filter plate:
1
Clean the filter plate mechanically.
2
Clean the filter plate chemically.
3
Rinse the filter plate under running tap water.
4
Protect the filter plate against drying out by storing it in the Filter-
CL cleaning container or in a closed plastic bag.
NOTE
Protect the filter plate from drying out. If the filter plate dries out,
it becomes hydrophobic, blocked and no more sample solution
can go through.
If the filter plate has dried out anyway, proceed as follows:
Submerse the filter plate in a 30 % ethanol or spirit solution for about
60 seconds (approx. 5 l solution are required for the filter plate to be
completely submersed in the Filter-CL cleaning container).
This reduces the surface tension of the filter plate, and the flow rate
should come up to the usual value. For fault-free operation, the filter
Maintenance / cleaning