Xylem YSI FM/PC Mounting Instructions

Filter module for analyzers

Advertisement

Quick Links

ba76171e02 04/2019
FM/PC
FM-Case/PC, Filter/PC
FILTER MODULE FOR ANALYZERS

Advertisement

Table of Contents
loading

Summary of Contents for Xylem YSI FM/PC

  • Page 1 ba76171e02 04/2019 FM/PC FM-Case/PC, Filter/PC FILTER MODULE FOR ANALYZERS...
  • Page 2 FM/PC, FM-Case/PC, Filter/PC For the most recent version of the manual, please visit www.ysi.com. Contact 1725 Brannum Lane Yellow Springs, OH 45387 USA Tel: +1 937-767-7241 800-765-4974 Email: info@ysi.com Internet: www.ysi.com © 2019 Xylem Inc. Copyright ba76171e02 04/2019...
  • Page 3: Table Of Contents

    FM/PC, FM-Case/PC, Filter/PC Overview ........5 Scopes of delivery .
  • Page 4 FM/PC, FM-Case/PC, Filter/PC ba76171e02 04/2019...
  • Page 5: Overview

    FM/PC, FM-Case/PC, Filter/PC Overview Overview For failure-free operation of the analyzer, the samples have to be sup- plied in a suitable quality and in sufficient quantities. The filter module is part of the system for sample filtration for the follow- ing analyzers: ...
  • Page 6: Mounting / Dismantling Of The Filter

    Mounting / dismantling of the filter module FM/PC, FM-Case/PC, Filter/PC Mounting / dismantling of the filter module Bild 3-1 Mounting of frame and filter plate 1 Frame parts (FM-Case/PC) 2 Adapter 3 Allen screws (hexagon socket head cap screws) 4 Filter plate Mounting Place the lower frame part (1) on a clean, flat surface.
  • Page 7: Installing The Filter Module

    FM/PC, FM-Case/PC, Filter/PC Installing the filter module The frame only serves as a bracket, not as a seal. The fil- ter plate is tight itself. NOTE Make sure that the frame is clean for the mounting. If solid parti- cles such as sand grains are between the frame and filter plate, the membrane could be damaged when the frame parts are screwed together.
  • Page 8: Maintenance / Cleaning

    Maintenance / cleaning FM/PC, FM-Case/PC, Filter/PC Further proceeding as described in the operating manual of your analyzer (chapter C OMMISSIONING Bild 4-2 Filtration unit (filter module mounted on the guide slide) Maintenance / cleaning Cleaning the filter plate When to clean the filter plate can be determined empirically based on the negative pressure display on the filtration pump.
  • Page 9: Mechanical Cleaning

    FM/PC, FM-Case/PC, Filter/PC Maintenance / cleaning approx. -0.1 bar. If the negative pressure increases by a further -0.3 ... -0.4 bar with time, the filter plate is covered with solid matter from the sample and has to be cleaned.  When the contamination of the filter plate grows, the negative pres- sure in the non-active condition of the filtration pump increases.
  • Page 10: Chemical Cleaning

    Maintenance / cleaning FM/PC, FM-Case/PC, Filter/PC • Do not dismantle the filter plate (Filter/PC) from the frame (FM- Case/PC). Rinse off any coarse contamination with low water pressure (e.g. with a watering can or wash bottle). Then carefully remove the coating from the filter plate using the soft special brush (scope of delivery FM-Case/PC).
  • Page 11 FM/PC, FM-Case/PC, Filter/PC Maintenance / cleaning In the event of calcification, diluted hydrochloric acid (HCl) is recom- mended (0.01 mol/l HCl, 0.036%, pH 2). Preparing the cleaning The cleaning solution for the filter plate can be made from different solution detergent concentrates by diluting with water.
  • Page 12: Exchanging The Filter Plate

    Maintenance / cleaning FM/PC, FM-Case/PC, Filter/PC Monitor the success of the cleaning and extend the reaction time as necessary. After the chemical cleaning, rinse the filter module with clean water. Even filter plates that are supposedly irreversibly blocked may be restored by leaving them in the cleaning solution for a longer period of time and then rinsing them several times.
  • Page 13: Storing A Used Filter Plate

    FM/PC, FM-Case/PC, Filter/PC Maintenance / cleaning Clean the filtration unit. Dismount the filter module from the guide slide. Unscrew the Allen screws of the frame (6 hexagon socket screws). Remove the used filter plate with the adapter. Clean the frame parts. Remount the new filter plate in the frame (see section and section OUNTING...
  • Page 14: Technical Data

    Technical data FM/PC, FM-Case/PC, Filter/PC plate must not dry out again, otherwise the procedure has to be repeat- Technical data Frame L 355.5 mm x W 266.5 mm x D 27.5 mm Dimensions Frame Materials Screws Stainless steel V4A Adapter Stainless steel V4A, PVDF, PVC ba76171e02 04/2019...
  • Page 15: Contact Information

    FM/PC, FM-Case/PC, Filter/PC Contact Information Contact Information Ordering & Technical Support Telephone: (800) 897-4151 (937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET Fax: (937) 767-1058 Email: info@ysi.com Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 Internet: www.ysi.com When placing an order please have the following information available:...
  • Page 16 Contact Information FM/PC, FM-Case/PC, Filter/PC ba76171e02 04/2019...
  • Page 18 150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation. For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241;...

This manual is also suitable for:

Ysi fm-case/pcYsi filter/pc

Table of Contents