SICK TRANSIC111LP Operating Instructions Manual

SICK TRANSIC111LP Operating Instructions Manual

Laser oxygen transmitter
Table of Contents

Advertisement

Quick Links

Title Page
MMMOPERATING INSTRUCTIONS
MMMI
O P E R A T I N G I N S T R U C T I O N S
TRANSIC111LP
Laser Oxygen Transmitter

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TRANSIC111LP and is the answer not in the manual?

Questions and answers

Summary of Contents for SICK TRANSIC111LP

  • Page 1 Title Page MMMOPERATING INSTRUCTIONS MMMI O P E R A T I N G I N S T R U C T I O N S TRANSIC111LP Laser Oxygen Transmitter...
  • Page 2 Subject to change without notice. © SICK AG. All rights reserved. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 3 Hazard or unsafe practice which could result in less severe or minor injuries. NOTICE Hazard which could result in property damage. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 4: Table Of Contents

    Connecting the TRANSIC111LP via an 8-pole plug connector... 29 3.4.4 Test gas connection (optional) ............ 30 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 5 Serial interface commands................40 4.6.1 List of serial interface commands ..........40 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 6 4.12.1 Set test current for analog output (command ITEST) ....53 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 7 Compensation of ambient parameters ............61 5.1.1 Pressure compensation..............62 5.1.2 Background gas effects..............62 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 8 One-point adjustment using the keypad (function CAL1)... 83 6.4.10 Restoring factory calibration............83 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 9 11.3 Type code TRANSIC111LP ............... 106 11.4 Password ....................107 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 10: Important Information

    Disconnect all connections to the power supply. ▸ Make sure accidental startup is not possible. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 11: Installation Location

    NOTICE: The TRANSIC111LP is not suitable for use in potentially explosive atmospheres. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 12: Responsibility Of User

    ▸ Keep passwords in a separate, safe place and secure against unauthorized use. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 13: Product Description

    Further information on components having contact with the sample gas probe, see “Dimen- sions and mechanics”, page 102. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 14: Transic111Lp Variants

    Gas feed line Outlet (back to process) from process O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 15: Variant For Ambient Gas Measurement

    . The measuring function is lost for oxygen concentrations under 2 vol%. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 16: Installation

    Further information on the operating temperature range can be found in “Ambient condi- tions”, page 101. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 17: Strong Light Sources Near The Oxygen Measuring Probe

    15 see “Dimensions and mechanics”, page 102. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 18: Installation

    Installation information: Process, materials and tools must be compatible with oxygen. Observe all regulations valid for your application for handling oxygen. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 19: Prerequisites For Assembly

    2 = Install the probe aligned horizontally or at a maximum downward angle of 45 ° in high humidity (5° downwards is recommended). 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 20: Installation Options

    DN50 flange (fitted with M16 hex bolts). The flange adapter is fitted at the factory and is fastened to the bottom of the adapter with a screw. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 21 “Adapter flange, weldable, suit- able for PS = 10 bar”, page 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 22: Transic111Lp Installation - Extractive

    TRANSIC111LP on the wall bracket. Tighten all four screws. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 23 This provides a better gas exchange and shorter reaction times. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 24 The Swagelok connections must point directly downwards. Fig. 9: Removing the sample gas cell O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 25: Transic111Lp Installation - Ambient Gas Measurements

    Ensure the TRANSIC111LP is installed in a representative gas mixture. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 26: Connections

    Fit the disconnecting device as close as possible to the measuring device and easily accessible. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 27 NOTICE: No potential differences may occur. NOTICE: Check the grounding regularly. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 28 13 Ensure the enclosure is closed tight. 14 The transmitter is now ready for use. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 29: Connecting The 24V Pelv Power Supply Unit

    Alarm green Table 1: Connections of 8-pole plug connector 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 30: Test Gas Connection (Optional)

    The test gas connection has a non-return valve with opening pressure of 1.7 bar (see “Connections and systems”, page 67). O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 31: Operation

    5 Back - Back button Back 6 Ent - Input button 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 32: Maintenance Interface

    The maintenance interface provides more configuration options for alarm threshold(s) or other settings than the keypad and display. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 33: Rs-485 Interface

    These settings can be changed via the maintenance interface. The contact relay is momentary. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 34: Settings Using The Keypad

    Unauthorized changing of parameters can have severe consequences. This is why the password is only available for authorized technicians. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 35: Menu Navigation Without Password

    Fig. 18: Displaying and resetting the temperature statistics Back Back Back Back 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 36: Calibration Gas, Actual Value (Cal.c)

    Fig. 21: Displaying all errors currently existing Back Back Back Back O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 37: Entering The Password (Pas)

    Fig. 24: Humidity in process gas setting (0 ... 600 g/m³) Back Back 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 38: Content In Process Gas: Settings (Co2)

    Fig. 27: Resetting to the factory setting for oxygen measurement < Back Back O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 39: Scaling The Analog Output (Ascl)

    Fig. 30: Restarting the oxygen sensor TRANSIC111LP Back Back 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 40: Serial Interface Commands

    Display/set serial communication settings Table 5: List of serial interface commands without password O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 41 Table 6: List of additional serial interface commands with password 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 42: Output Of Measuring Results

    Character string specifying the output format for the measuring results Example: >send 20.9 20.8 24.5 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 43: Display/Set Serial Communication Mode (Command Smode)

    = Serial communication mode, possible modes are STOP, POLL and RUN Example: >smode2 SMODE : STOP ? > 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 44: Display/Define Serial Communication Settings (Command Seri)

    : 19200 ? DATA BITS PARITY : NONE ? STOP BITS O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 45: Display Status Of Serial Maintenance Interface (Command Sci1)

    The formatting string comprises the data to be shown and the associated formatting commands. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 46 The standard output format is used with command FORM /: >form / >send Oxygen = 21.0 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 47: Display/Set Date (Command Date)

    = Seconds Examples: >time 03:28:32 >time 11:23:01 11:23:01 > 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 48: Network Operation

    Syntax: CLOSE<cr> Example: >close line closed O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 49: Set Echo Mode (Command Echo)

    TRANSIC100LP 9165087 0000 / 1.36 echo on ECHO : ON 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 50: Commands For Access At Maintenance Level

    = Water content in measured gas (g/m 0511-086 Example: >H2O 100 WATER(g/m3) : 100 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 51: Set Carbon Dioxide Content For Compensation (Command Co2)

    Ilow (mA) ? 3.42 Ihigh (mA) ? 17.6 > 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 52: Scaling And Setting The Analog Output

    Pressure = Pressure setting (bar(a)) Example: >xpres 1.300 PRESSURE(bar) : 1.300 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 53: Checking The Analog Output

    LO POINT (%02) : 10.0 HI POINT (%02) : 11.0 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 54: Device Information And Other General Commands

    - Gas temperature (centigrade) TGASF - Gas temperature (fahrenheit) O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 55: Display Calibration Information (Command Cinfo)

    Prints information about the device Prints information even in POLL mode > 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 56: Display Status Of Laser Temperature Controller (Command Ltc).56

    : 3.000 OUTMAX02 (%) : 20.000 OUTMIN02 (%) : 0.000 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 57: Display All Changeable Parameter Values (Command Param)

    : 23.39 Ti max : 32.53 Ti min : 24.55 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 58: Display Status Of Submenu Item (Command Status)

    EEPROM (op) saved successfully EEPROM (op_log1) saved successfully EEPROM (op_log2) saved successfully O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 59: Resetting The Measuring Device

    : WARNING ERRORS: WARNING : WATCHDOG RESET OCCURRED > 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 60: Display Errors Detected (Command Errs)

    0:NON FATALSIGNAL CUT 52:OFF: 0:WARNINGEEPROM LOG&STATS CORRUPTED 53:OFF: 0:WARNINGWATCHDOG RESET OCCURRED O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 61: Setting Ambient Parameters

    = 1.2 bar --- not compensated Process pressure (bar(a)) 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 62: Pressure Compensation

    (factory setting). O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 63 Collisions between the water and oxygen molecules affect the shape of the oxygen absorption lines. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 64 “Influence of background gases on oxygen measurement”, page 105. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 65: Adjustment

    “Calibration”, page 79, and/or see “Infor- mation on adjustment”, page 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 66: Setting Up The Gas Supply For Calibration And Adjustment

    Ensure that escaping gas is safely discharged. Fig. 33: Fastening the TRANSIC111LP probe in the sample gas cell O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 67: Calibration And Adjustment In The Process

    Although the non-return valve prevents process gas from escaping, the plug on the optional TRANSIC111LP gas inlet can also be used. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 68: Adjusting Gas Flow

    Allow enough time for the gas concentration to stabilize when doing calibrations/ adjustments. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 69: Calibration

    10.4 Table 11: Oxygen measured values for relative humidity 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 70: Using Bottled Gas

    Ensure the analog output is no longer held. ● O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 71: Adjustment

    Two-point adjustment: Always returns new gain and offset values. ● 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 72: One-Point Adjustment Via The Serial Interface (Command Coxy1)

    EEPROM (op) saved successfully EEPROM (op_log1) saved successfully EEPROM (op_log2) saved successfully O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 73: One-Point Adjustment Using The Keypad (Function Cal1)

    – Change to the gain value: Oxygen concentration > 10.5% O 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 74: Two-Point Adjustment Via The Serial Interface (Command Coxy2)

    Enter command COXY1 and confirm with the Enter button. Connect the gas inlet and let the gas flow in. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 75: Connect Ref Gas #2 To Cuvette

    >adjust off Outputs to normal state 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 76: Two-Point Adjustment Using The Keypad (Function Cal2)

    Fig. 36: Two-point adjustment using the keypad Back Back < Back Back < O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 77 – Change to the gain value: Oxygen concentration > 10.5% O 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 78: Transic111Lp Adjustment For Ambient Gas Measurement

    (20.95 O O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 79: Calibration

    Allow enough time for the gas concentration to stabilize when doing calibrations/ adjustments. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 80: Calibration Method

    21.0%) via the keypad. – Restoring factory calibration O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 81: One-Point Adjustment Via The Serial Interface

    Connect the gas inlet and let the gas flow in. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 82 The adjustment is finished and the output returns to displaying the measurement results. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 83: One-Point Adjustment Using The Keypad (Function Cal1)

    59, via the keypad, see “Resetting the measur- ing device (rESE)”, page 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 84: Maintenance

    ▸ Only open the TRANSIC111LP in a dry environment free from dust. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 85: Cleaning The Optical Components

    If the surface is still contaminated, tip pure ethanol or isopropanol onto the surfaces. Let the chemicals work in for maximum 15 minutes. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 86 Cleaning the TRANSIC111LP optics is easiest with an optics cleaning set from SICK. Part No., see “Accessories”, page O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 87: Cleaning The Transic111Lp Filter

    3 + 4 Areas that can be touched by hand 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 88: Cleaning The Filter

    ▸ Make sure to rinse the optical components thoroughly when using cleaning agents. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 89: Spare Parts And Accessories

    Screw fitting M20*1.5 D8-11 2060180 Screw fitting, M20*1.5 on 1/2"NPTf CUZN 2060179 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 90 Connection line, plug (M12), 2 m 2059455 Plug M12 2060101 Optics, cleaning set 2072979 O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 91: Troubleshooting

    Analog output 0.0 mA Red LED on Digital output Open 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 92: Error Display

    Supply voltage outside the allowed range. Check (non-serious) voltage supply/cables. Table 13: Error Table O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 93 WATCHDOG RESET OCCURRED Reset through software error. Table 13: Error Table 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 94: Shutdown

    CAUTION: Hazard through process residues on the device ▸ Clean the TRANSIC111LP before shipping to SICK. O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 95: Specifications

    A 24V PELV power supply unit is mandatory for supplying the TRANSIC111LP to ensure electrical safety. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 96: Technical Data

    All specifications in mm (inches) 1 = ø 6.5 mm, four pieces O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 97 16,1 (0.63) 38,8 (1.53) All specifications in mm (inches) 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 98 Fig. 46: Adapter flange, weldable, suitable for PS = 10 bar All specifications in mm (inches) O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 99 (3.54) (8.27) (4.06) (1.81) All specifications in mm (inches) 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 100 (13.86) (8.15) (1.97) (3.94) All specifications in mm (inches) O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 101: Measured Value Recording

    Use of TRANSIC111LP in potentially explosive atmospheres is not permitted. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 102: Transic111Lp Inputs And Outputs

    ● PTFE, SiN, MgF2, quartz ● Polymer coating ● O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 103: Pressure Suitability

    Test gas connection Swagelok connection for 6 mm tubes 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 104: Annex

    11.1 Humidity Conversion Table Table 14: Humidity Conversion Table O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK Subject to change without notice...
  • Page 105: Influence Of Background Gases On Oxygen Measurement

    If you have questions concerning further background gases not listed here, please ● contact Sick Service. 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 106: Type Code Transic111Lp

    Example: TRANSIC111LP- A 2 B 1 A 1 A 1 A 1 A 1 A O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP 8014084/ZWT7/V3-1/2020-06 | SICK...
  • Page 107: Password

    ANNEX 11.4 Password 1010 8014084/ZWT7/V3-1/2020-06 | SICK O P E R A T I N G I N S T R U C T I O N S | TRANSIC111LP Subject to change without notice...
  • Page 108 Phone +7 495 283 09 90 E-Mail info@sick.com E-Mail office@sick.com.gr E-Mail info@sick.ru Vietnam Hong Kong Singapore Phone +65 6744 3732 Phone +852 2153 6300 Phone +65 6744 3732 E-Mail sales.gsg@sick.com E-Mail ghk@sick.com.hk E-Mail sales.gsg@sick.com Detailed addresses and further locations at www.sick.com...