Table 9 Gmw80 Compliance; Table 10 Gmw80 Operating Environment - Vaisala CARBOCAP GMW80 Series User Manual

Carbon dioxide, humidity, and temperature transmitters
Hide thumbs Also See for CARBOCAP GMW80 Series:
Table of Contents

Advertisement

GMW80 Series User Guide
Property
Response time (63 %)
Carbon dioxide sensor
Temperature
Measurement range
Temperature sensor
Accuracy (GMW83 and GMW84):
at +10 ... +30 °C (+50 ... +86 °F)
at +0 ... +10 °C (+32 ... +50 °F) and +30 ... 50 °C
(+86 ... +122 °F)
Humidity
Measurement range
Accuracy at temperature range +10 ... +30 °C (+50 ... +86 °F):
0 ... 80 %RH
80 ... 95 %RH
Accuracy at temperature ranges 0 ... +10 °C (+32 ... +50 °F) and +30 ... +50 °C (+86 ... +122 °F):
0 ... 95 %RH
Stability in typical HVAC applications
Product lifetime
Table 9 GMW80 compliance
Property
EMC compatibility
Compliance marks
Table 10 GMW80 operating environment
Property
Operating temperature
Operating humidity
44
Value/Description
60 s
7 min (GMW87 and GMW88)
Vaisala CARBOCAPâ GM10
0 ... +50 °C (+32 ... +122 °F)
On P models: Pt1000 RTD Class F0.15 IEC 60751
For analog outputs: Digital temperature sensor
±0.5 °C (0.9 °F)
±1 °C (1.8 °F)
0 ... 95 %RH
±3 %RH
±5 %RH
±7 %RH
±2 %RH over 2 years
> 15 years
Value/Description
IEC/EN/BS 61326‑1, industrial environment
CE, RCM, WEEE
Value/Description
0 ... +50 °C (+32 ... +122 °F)
0 ... 95 %RH
Dew point < 30 °C (+86 °F)