Leds - O-Two e700 User Manual

Automatic transport ventilator
Table of Contents

Advertisement

At approximately 2% of full battery capacity, the ventilator will not
start when it is turned o or will shut down when it is turned on.
Note: The battery capacity level is detected from measured voltages
and the capacities shown above are based on results from new
batteries tested at room and low temperature. Battery capacity levels
are subject to change when old batteries are used.
Fully charged Batteries shall be recharged after 6 months without
usage or when battery discharging LED turns orange and fl ashes.
The battery re-charge time is about 5.5 hours from fully discharged.
Batteries have a minimum of 200 discharge and charge cycles.

7.3. LEDs

46
WARNING
Green color LED - Continuous when unit is ON and fl ashing
when unit is OFF.
Red or Yellow color LED – Flashing during alarm/ warning situation.
LED color depends on the severity of the failure.
Green color LED - Continuous when unit is connected to AC
power source during both ON and OFF phases.
Orange color LED - Continuous when unit is charging and o when
battery is fully charged during both ON and OFF phases.
During OFF phase this light will start fl ashing when battery capacity
drops to around 90 %.
Green color LED - Continuous when unit is operated
using internal battery.
e700 User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

01eve700

Table of Contents