Figure 13 - CPAP ventilation waveform

5.2.5. CPR mode

The CPR mode consists of timed chest compression audible prompts
coupled with automatically delivered breaths for both intubated
and mask ventilated patients. There is also a visual animated display
to guide the health care provider while performing CPR.
The CPR mode for masked ventilated patients is the default setting
for this mode but changes can be made between the 2 sub-modes
at any time.
The CPR mode for masked ventilated patients consists of 2 phases,
chest compression and ventilation. 30 chest compressions over
18 seconds are synchronized with audible prompts and on screen
visual animations, followed by two, 1 second, mandatory breaths
within a 5 second ventilation phase. The ratio between chest
compressions and ventilations is 30:2.
The CPR mode for intubated patients consists of continuous
compressions indicated by an audible prompt and visual animation
at a rate of 100 compressions per minute plus automatically
delivered breath every 6 seconds (10 BPM).
The default ventilation in CPR mode is fl ow controlled ventilation.
The default tidal volume is set according to the initial start-up
39