5.2.1. A/C V (Assist Control Ventilation)

In this mode the ventilator can deliver either volume ventilation
(VCV) if Tidal Volume (Vt) is selected (Figure 10.a) or pressure
ventilation (Figure 10.b) if Pressure Controlled Ventilation (PCV) is
selected. Choosing either mode will disable the other one which
will be displayed with (-) on the screen.
Volume control ventilation (VCV) with Tidal Volume and ventilation
Rate setting according to patient size is the default start up for this
mode (see Table-2 below).
During A/C V mode, the ventilator will deliver Controlled Mandatory
Ventilation (CMV) regardless of any patient's e ort if the trigger
(Trig.) is disabled (displayed with (-)).
The default trigger for A/C V is 3 L/min but can be adjusted up to
15 L/min.
If no inspiratory e ort is detected during the trigger window, the
ventilator will initiate mandatory ventilation at the end of trigger
window. Should the patient demand more fl ow than set by user, he/
she can withdraw the excess from ambient.
TABLE 2 - Default Ventilation Setting- A/C V
PARAMETER
Tidal Volume
Rate
Mv
I:E ratio
Ti*
PEEP
PCV
Trig.
O2 %
Mv max
Mv min
P max.
P min.
Manual
e700 User Manual
RANGE
(50 - 2000 ml)
(5 - 60 BPM)
Calculated based on
Vt & f
(1:4 – 3:1)
(0.2 – 9 sec.)
(OFF, 4-20 cm H2O)
(OFF, 4-50 cm H2O)
(OFF, 1 -15 L/min)
(100% or 60% O2)
(2 -40 L)
(0.5 – 35 L)
10 -80 cm H2O
0-20 cm H2O during I
time only
Refer to Manual and
I-Hold section
DEFAULT
INFANT
CHILD
100
30
3.0
1:2
0.66
5
O , unless
O , unless
selected
selected
3
100
30
0.5
25
3
ready
ready
ADULT
250
500
15
10
3.7
5.0
1:2
1:2
1.33
2.0
5
5
O , unless
selected
3
3
100
100
30
30
0.5
0.5
25
30
3
3
ready
31