Notes On Maintenance; Camera Electronic; Optical System; Cleaning The Infra-Red Barrier Filter - Zeiss Axiocam MR Installation Reference

Mobile stand-alone color 5 megapixel microscope camera
Hide thumbs Also See for Axiocam MR:
Table of Contents

Advertisement

NKR= kçíÉë=çå=j~áåíÉå~åÅÉ=
`~ãÉê~=bäÉÅíêçåáÅ=
Camera and interface are maintenance free. If necessary, cleaning of the
camera can be performed with normal microscope cleaning material, but do
never use compressed air or other
léíáÅ~ä=póëíÉã=
The internal optical components of the camera should always be protected. If
no lens or TV adapter incorporating optics is screwed into the camera's C-
Mount thread, the camera's sensor and IR-filter must be protected by
screwing the protective cap into the camera's C-Mount thread.
`äÉ~åáåÖ=íÜÉ=fåÑê~JêÉÇ=_~êêáÉê=cáäíÉê=
Contamination of the infrared filter has an adverse effect on the quality of the
resulting image (dark points, cloudy structures in the image). If there is dry
dust on the front side of the infrared filter, you can remove it with a soft
brush or with cotton (wool) after unscrewing the lens or TV adapter.
fãéçêí~åí=kçíÉ=çå=íÜÉ=rëÉ=çÑ=`Jjçìåí=
lÄàÉÅíáîÉë=
The camera can be mounted onto microscopes with standard TV-adapters
with C-mount connections. Carl Zeiss supplies the camera complete with a
pre-mounted infrared barrier filter, which offers the optical advantage for use
on microscopes of a lower sensitivity to dust.
The IR barrier filter is situated 5 mm behind the outer edge of the C-mount
opening in the camera.
Due to this pre-mounted IR barrier filter, C-mount objectives that
screw more than 5 mm into the thread cannot be screwed into the
camera.
AxioCam MR Installation + Reference
Installation of the AxioCam MR
23