Problem
Possible Cause
The ambient temperature
is lower than 10°C (50.0°F)
"Er1" is displayed.
or higher than 40°C
(104°F).
The lens of the temperature
probe is dirty.
The temperature
The thermometer probe is not
reading is lower
aligned to the eardrum.
than the typical
body temperature
The thermometer is used
range.
within 30 minutes after being
taken from a cold environ-
ment.
The temperature
reading is higher
The temperature probe is
than the typical
damaged.
body temperature
range.

Specifications

Product Name
Infrared Thermometer
Product Model
JPD-FR412
Power Supply
Internal power supply
Mode
Operating Voltage
DC 3V
Battery Model
AAA x 2
Operating Mode
Continuous operating
Display
Segment LCD
Measure time
About 1 second
Latency Time
About 3 seconds
Forehead mode:35.0°C–42.2°C (95.0°F–108.0°F)
Measuring Range
Ear mode:35.0°C–42.0°C (95.0°F–107.6°F)
Object mode:0.0°C–100.0°C (32.0°F–212.0°F)
Solution
Take a measurement
under an ambient tem-
perature between 10°C
(50.0°F) and 40°C (104°F).
Clean the lens using a cotton
swab.
Reposition the thermometer
probe so that it is aligned to
the eardrum.
Wait for more than 30 minutes
after the thermometer is
moved into the measurement
environment.
Contact the manufacturer.
Accuracy
Forehead mode: ±0.2°C (±0.4°F)
(Laboratory)
Ear mode: ±0.2°C (±0.4°F)
Object mode: ±1.0°C/±2.0°F
Resolution
0.1°C (0.1°F)
Memory
20 temperature readings
The low-battery symbol is displayed if the power
Low-battery Alert
voltage is lower than 2.4 V±0.1V
Automatic
The thermometer automatically powers off if it is not
Power-off
used in 10±1 seconds.
Outer
155.9*40.2*49.2mm
dimensions (mm)
Weight (g)
Thermometer (with batteries): 90g
Temperature: 10°C~ 40°C (50°F–104°F)
Operating Envi-
Humidity: 15%–95% RH, non-condensing
ronment
Atmospheric pressure: 86–106 kPa
The infrared thermometer has been tested and conforms to the standard
ASTM E1965-98. ASTM laboratory accuracy requirements in the display range
of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is ±0.4°F
(±0.2°C). Note that for mercury-in-glass and electronic thermometers, the
requirement per ASTM Standards E667-86 and E1112-86 is ±0.2°F (±0.1°C).

Security Class

Type of protection against electric shock: internally powered equipment.
Degree of protection against electric shock:
- Degree of protection against ingress of water:IP22
- Safety degree of using in flammable anesthetic gas blending with air, oxygen
or nitrous oxide: Non-AP/APG
- No applied parts of the thermometer prevents defibrillation charge effect.
- No applied parts of the thermometer output signal.
- The thermometer is impermanent installed device.

Storage and Transportation

The thermometer can be transported using general transportation tools.
Severe vibration, shock, or rain must be avoided during transportation.
The thermometer must be packaged and then stored in a well-ventilated room
without corrosive gas. The ambient temperature must be between –20°C
and +55°C -4°F–131°F , the relative humidity must be lower than 95% (non-
condensing), and the atmospheric pressure must be 50–106 kPa.
Type BF applied part.