4. Reprocessing Procedure

The probes must be reprocessed after each use. Refer to the
reprocessing instruction in this chapter. For reprocessing of the
available options for the probes, please refer to the specific
instruction manuals of them.
WARNINGS
Limitations on
reprocessing
Transportation
before using
The probes are delivered unsterile. Prior to the
first use, reprocess the probes.
Temperature should not exceed 60°C during
reprocessing.
Probe connector is not water resistant.
The probe is not completely submersible. The
immersible part is shown in Fig.1. The un-immersible
part should be disinfected by wipe disinfection.
The probe should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful
not to damage the sterile pouch or container during
transportation.