Information contained in this document is the property of Nüve. It may not be duplicated or distributed without its permission. NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Saracalar Mahallesi, Saracalar Kümeevleri No:4/2 Akyurt 06750 ANKARA Ankara / TURKEY : (90.312) 399 28 30 (pbx) : (90.312) 399 21 97 Sales : sales@nuve.com.tr Technical Service: service@nuve.com.tr...
Page 3
INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NÜVE. IT MAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT PERMISSION. PLEASE REGISTER ONLINE TO VALIDATE YOUR WARRANTY: To register your warranty online, please visit our webpage www.nuve.com.tr and fill in the “Warranty Registration Form”.
COMPLETION OF THE WORK PERIODIC MAINTENANCE AND CLEANING 7.1. PERIODIC MAINTENANCE 7.2. CLEANING DISPOSAL MANAGEMENT CONCEPT TROUBLESHOOTING 9.1. ERROR CODES 9.2. FUSE REPLACEMENT ELECTRICAL CIRCUIT DIAGRAM 10.1. ELECTRICAL CIRCUIT DIAGRAM FOR EN 032 10.2. ELECTRICAL CIRCUIT DIAGRAM FOR EN 055 AND EN 120...
1. INTRODUCTION 1.1. USE AND FUNCTION The EN 032 incubator is designed to provide excellent incubation conditions for applications in biology and microbiology laboratories such as medical and veterinary fields; research and quality control examinations in pharmaceutical, food and cosmetics industries and biotechnology.
Shelf carrier (EN 032) K 23 048 Shelf carrier (EN 055) K 23 040 Shelf carrier (EN 120) R 01 132 Mesh type shelf (EN 032) R 01 127 Mesh type shelf (EN 055) R 01 031 Mesh type shelf (EN 120) ...
The non-adherence to these warnings can lead to material damage and injury to personal. Symbol in the operating instructions: This symbol refers to important circumstances. Labels on the device: EN 032 Fuses EN 055 Fuses EN 120 Fuses Topraklı Surface Priz...
5. INSTALLATION 5.1. ENVIRONMENTAL CONDITIONS The instrument is design to operate safely under the following conditions: Indoor use only Ambient temperature: 5°C to 40°C Maximum relative humidity for temperature up to 22°C: 80% Maximum altitude: 2000 m ...
Leave at least 20 cm free space between the equipment and wall. 5.6. GENERAL PRESENTATION VENTILATION ADJUSTMENT BUTTON DISPLAY AND CONTROL PANEL DOOR HANDLE PLASTIC PADS Figure 1 SAFETY THERMOSTAT ON / OFF SWITCH FUSE HOLDER AND SUPPLY INLET VENTILATION HOLE...
Figure 2 5.7. CONTROL PANEL Figure 3 1- Heating Process Display: This display shows the chamber temperature during the operation and the set temperature value during programming. The error codes are also shown on this display. 2- Time Display: This display shows the elapsed time during the operation and the set time value during programming.
At most 70% of the surface area of the shelves should be used in order to obtain a uniform temperature distribution. 6. OPERATING PRINCIPLES 6.1. PROGRAMING Push the temperature set key. Set the temperature by pushing the value increase and decrease keys on the temperature adjustment side.
Before operation starts, check that the safety thermostat value is higher than the temperature set value. 6.2. COMPLETION OF THE WORK Observe that the program is over. Stop the operation by pushing the “start/stop” button. Take the samples out. ...
The instruments and electronic accessories (without batteries, power packs etc.) must be disposed off according to the regulations for the disposal of electronic components. Batteries, power packs and similar power source have to be dismounted from electric/electronic parts and disposed off in accordance with applicable local regulations. 9.
10. ELECTRICAL CIRCUIT DIAGRAM 10.1. ELECTRICAL CIRCUIT DIAGRAM FOR EN 032 10.2. ELECTRICAL CIRCUIT DIAGRAM FOR EN 055 AND EN 120 HEATERS: EN 055 2 x 700 W EN 120 2 x 800 W...